Operating instructions

Special applications and functions
52
The following displays appear during the adjustment (calibration):
The internal adjustment weight is being loaded.
The internal adjustment weight is being raised.
The balance is processing the adjustment results.
The balance reports successful completion of the adjustment (calibration).
The balance automatically returns to the weighing mode.
You can always abort an ongoing adjustment (calibration) by briefly
pressing the «C» key.
If the adjustment (calibration) can not be performed properly (e.g. as a result
of vibrations), the balance aborts the adjustment operation and “Abort”
appears in the display. Press the «C» key to clear this message and restart
the adjustment operation.
If your balance is connected to a printer, the adjustment (calibration) is
recorded automatically in conformance with GLP. The record shown oppo-
site is a specimen printed with the METTLER TOLEDO LC-P45 Printer.
Depending on the attached printer, the printout may differ somewhat from the
example shown.
Cal
------
Cal
=00
Cal
------
CAL donE
=0002
g
C
Abort
--BALANCE CALIBRATION--
03.02.97 11:23:34
METTLER TOLEDO
Balance
Type: AG204DR
SNR: 23001222
Int. calibration done
Signature:
........................
--------- END ----------