Operating instructions

Special applications and functions
54
AG 31
Place the requested weight in the middle of the weighing pan.
During the calibration, the horizontal segments are displayed.
Note
You can abort the ongoing calibration at any time by briefly pressing the «C»
key.
On completion of the calibration procedure, you are prompted to lift off the
weight. Remove the weight from the weighing pan.
After removal of the weight, the balance shows the end of the calibration
procedure and then returns to the weighing mode.
Note
If the calibration can not be performed properly (e.g. owing to vibrations),
the balance aborts the calibration procedure and “Abort” appears in the
display. Press the «C» key to clear this message and restart the calibration
procedure.
If your balance is connected to a printer, the adjustment (calibration) is
recorded automatically in conformance with GLP. The record opposite is a
specimen printed out with the METTLER TOLEDO LC-P45 Printer. Records
printed with other printers may differ somewhat from the example shown.
CAL donE
Cal
------
=0000
g
Abort
C
--BALANCE CALIBRATION--
03.02.97 ,11:34:23
METTLER TOLEDO
Balance
Type: AG104
SNR: 54001222
Weight ID:..............
Weight: 100.0000 g
Ext. calibration done
Signature:
........................
--------- END ----------