Transmitter pH M300 V20 1 Operation Manual Transmitter pH M300 Transmitter pH M300 52 121 304 © 02/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter pH M300 52 121 304
Transmitter pH M300 V20 Revision History 1.0 Initial Version 1.1 Adjusted Screen shots to SW V1.27 1.11 Changed chapters 19, 20 to 18, 19 1.12 Changed title 11.1.1 1.13 Modified Chapter 2 1.14 Various smaller changes 1.15 Formatting 1.16 Adaption to Fieldtest inputs, Adjustments Spec 1.17 Adaption to HMI changes Finetuning layout 1.18 Adaption to HMI changes 1.19 Changed Wiring Examples, small changes to accommodate MMI Changes 1.
Transmitter pH M300 V20 3 Operation Manual Transmitter pH M300 © 02/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter pH M300 52 121 304
Transmitter pH M300 V20 4 Index 1 Introduction _______________________________________________________________________________ 7 2 Safety instructions __________________________________________________________________________ 7 2.1 Definition of equipment and documentation symbols and designations_________________________ 7 2.
Transmitter pH M300 7.6 7.7 7.8 8 Configure 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 V20 5 7.5.1 One-point temperature calibration ______________________________________________ 33 7.5.
Transmitter pH M300 11.3 V20 6 11.2.2 Calibrate analog____________________________________________________________ 59 11.2.3 Calibrate unlock ____________________________________________________________ 60 Tech Service _____________________________________________________________________ 60 12 Info _____________________________________________________________________________________ 61 12.1 Messages _______________________________________________________________________ 61 12.
Transmitter pH M300 1 V20 7 Introduction Statement of Intended Use - The M300 pH transmitter is a single channel online process instrument for measuring pH of fluids. It will interface with a variety of different Mettler-Toledo sensors, which connect to the transmitter using cables of varied lengths. A large four line backlit Liquid Crystal Display conveys measuring data and setup information. The menu structure allows the operator to modify all operational parameters by using keys on the front panel.
Transmitter pH M300 V20 8 WARNINGS: Installation of cable connections and servicing of this product require access to shock hazard voltage levels. Main power and relay contacts wired to separate power source must be disconnected before servicing. Main power must employ a switch or circuit breaker as the disconnecting device for the equipment. Electrical installation must be in accordance with the National Electrical Code and/or any other applicable national or local codes.
Transmitter pH M300 3 V20 9 Unit overview M300 models are available in both a 1/4DIN and 1/2DIN case size. The 1/4DIN is a panel-mount only design and the 1/2DIN models provides an integral NEMA 4X/IP65 housing for wall-, pipe- or panel-mount. 3.
Transmitter pH M300 3.
Transmitter pH M300 V20 3.3 Control/navigation keys 3.3.1 Menu structure 11 Below is the structure of the M300 menu tree: Measurement Mode Menu Quick Setup Configure System Measurement Channel Setup Temp.
Transmitter pH M300 3.3.2.2 Press the 3.3.2.3 Use the 3.3.2.4 Press the 3.3.2.5 Press the 3.3.2.6 Press the 3.3.3 V20 Escape and key simultaneously (escape) to return to the Measurement mode. Enter key to confirm action or selections. Menu key to access the main Menu. Calibration mode key to enter Calibration Mode. Info mode key to enter Info Mode Navigation of data entry fields Use the key to navigate forward or the fields of the display. 3.3.
Transmitter pH M300 3.3.5 V20 13 Navigation with ↑ in Display If a ↑ is displayed on the bottom right hand corner of the display, you can use the or the key to navigate to it. If you click [Enter] you will navigate backwards through the menu (go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu. 3.3.
Transmitter pH M300 V20 4 Installation instruction 4.1 Unpacking and inspection of equipment 14 Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Do not discard the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Mettler-Toledo immediately. 4.1.
Transmitter pH M300 4.1.2 - V20 15 Installation procedure - 1/4DIN models Make cutout in panel (see dimensions cutout drawing). Be sure surface surrounding cutout is clean, smooth and free of burrs. Slide face gasket (supplied with transmitter) around transmitter from the back of the unit. Place transmitter into cutout hole. Be sure there are no gaps between the transmitter and panel surface.
Transmitter pH M300 4.1.3 V20 16 Panel cutout dimensional information - 1/2DIN models 1/2DIN Model transmitters are designed with an integral rear cover for stand-alone wall mount installation. The unit may also be wall mounted using the integral rear cover. See installation instructions in Section 4.1.4. Below are cut-out dimensions required by the 1/2DIN models when mounted within a flat panel or on a flat enclosure door. This surface must be flat and smooth.
Transmitter pH M300 4.1.4 - V20 17 Installation procedure - 1/2DIN models Remove rear cover from front housing. Start by unscrewing the four screws located on the face of the transmitter, in each corner. This allows the front cover to swing away from the rear housing. Remove the hinge-pin by squeezing the pin from each end. This allows the front housing to be removed from the rear housing Drill out wall-mount breakthroughs in the rear housing.
Transmitter pH M300 4.2 V20 18 Connection of power supply All connections to the transmitter are made on the rear panel of all models. Be sure power to all wires is turned off before proceeding with the installation. High voltage may be present on the input power wires and relay wires. A two-terminal connector on the rear panel of all M300 models is provided for power connection. All M300 models are designed to operate from a 20-30 VDC or a 100 to 240 VAC power source.
Transmitter pH M300 4.2.
Transmitter pH M300 V20 20 4.3 Connector PIN definition 4.3.1 TB1 and TB2 for 1/2DIN and 1/4DIN versions 4.3.2 TB3 for 1/2DIN and 1/4DIN versions Sensor Colour Glass electrode Coax inner / transparent Reference electrode Outer shield/ Solution GND RTD return (T1) RTD in (T2) Pin no.
Transmitter pH M300 V20 4.4 Connection of sensor 4.4.1 Connecting the sensor to the VP cable 21 NOTE: Cable lengths > 20 m can worsen the response during pH measurement.
Transmitter pH M300 4.4.
Transmitter pH M300 V20 4.5 Wiring examples (using TB3) 4.5.1 Example 1 (using TB3) 23 pH measurement without Solution Ground NOTE: Jumper Terminal 3 and 4 Wire colors only valid for connection with VP cable, blue and grey not connected.
Transmitter pH M300 4.5.2 V20 24 Example 2 (using TB3) pH measurement with Solution Ground NOTE: Wire colors only valid for connection with VP cable, grey not connected.
Transmitter pH M300 4.5.
Transmitter pH M300 4.5.4 V20 26 Example 4 (using TB3) ORP measurement with pH Solution ground electrode (e.g. InPro 3250SG, InPro4800SG).
Transmitter pH M300 V20 27 5 Placing transmitter in, or out, of service 5.1 Placing transmitter in service After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is powered. 5.2 Placing transmitter out of service First disconnect the unit from the main power source, then disconnect all remaining electrical connections. Remove the unit from the wall / panel. Use the installation instruction in this manual as reference for disassembling mounting hardware.
Transmitter pH M300 6 V20 28 Quick setup (PATH: Menu/Quick Setup) While in Measurement mode press the [MENU] key to bring up the Menu selection. Select Quick Setup and press the [ENTER] key. The display prompts you to enter the Quick Setup security code. Enter the Quick Setup security code to “xxxxx”, then press the [ENTER] key to confirm the Quick Setup security code.
Transmitter pH M300 V20 29 Selecting Yes and pressing [ENTER] means a Setpoint can be configured for channel a. Following Setpoint Types can be selected: High, Low, Between (an alarm is triggered if the measured value is between the specified range), Outside (an alarm is triggered if the measured value is outside the specified range) After setting the Set point value(s) a Relay (none, 1, 2, 3, 4) can be configured for that Set Point. The Relay delay is set to 10 seconds and the Hysteresis is set to 5%.
Transmitter pH M300 7 V20 30 Calibration (PATH: Cal) 7.1 Exit calibration mode Note: To exit Calibration mode at any time press the and key simultaneously (escape). The transmitter returns to the Measurement mode and the old calibration values remain active. 7.2 Enter calibration mode While in Measurement mode press the key. The display prompts you to enter or key to set the calibration the calibration security code.
Transmitter pH M300 V20 31 Place the electrode in the buffer solution and press the [ENTER] key to start the calibration. Auto mode: The display shows the buffer that the transmitter has recognized (Point 1) and the measured value. Manual mode: Enter the buffer value and press [ENTER] to proceed. As soon as the drift conditions have been fullfilled (or [ENTER] pressed in manual mode) the display changes to show the slope calibration factor S and the offset calibration factor Z.
Transmitter pH M300 V20 32 Select Yes to save the calibration values and the successful Calibration is confirmed on the display. 7.3.3 Process calibration Select Process Calibration by pressing the once followed by the [ENTER] key. Take a sample and press the [ENTER] key again to store the current measuring Value. To show the ongoing Calibration Process an A is displayed in the top left hand corner. After determining the pH Value of the Sample press the key again to proceed with the calibration.
Transmitter pH M300 7.5 V20 33 Temperature calibration Enter Calibration mode as described in section 7.2. and select Temperature Calibration. 7.5.1 One-point temperature calibration See 7.5 on how to enter the Temperature Calibration menu. Slope or Offset can be selected with the 1 point calibration. Select Slope to recalculate the Slope factor M and Offset to recalculate the offset calibration factor A. Enter Point 1 The newly calculated value – either M or A – is displayed.
Transmitter pH M300 V20 34 The newly calculated values M and A are displayed. Select Yes to save the new calibration values and the successful Calibration is confirmed on the display. 7.6 Edit pH Enter Calibration mode as described in section 7.2. and select Edit pH. All calibration constants of the primary measurement (p) for the sensor are displayed on Line 3. All calibration constants of the secondary measurement (s) for the sensor are displayed on Line 4.
Transmitter pH M300 7.8 V20 35 Verify Enter Calibration mode as described in section 7.2. and select Verify. The value of the primary measurement (mV for a pH or ORP sensor) and the value of the secondary measurement (resistance of the temperature sensor for all sensors except ORP) are shown. The meter calibration factors are used when calculating these values.
Transmitter pH M300 8 V20 36 Configure (PATH: Menu/Configure) Configure Measurement Analog Outputs Set Points Channel Setup Temp. Source pH Set Averaging 8.1 Alarm/Clean/ Hold Display Hold Outputs Measurement Resolution Backlight Name Exit configuration mode NOTE: to exit Configuration mode at any time press the and key simultaneously (escape). The transmitter returns to the Measurement mode and the old settings remain active. 8.
Transmitter pH M300 8.3.1 V20 37 Channel setup Press the [ENTER] key to select this Menu. The 4 lines of the display can now be configured with a value. Pressing the [ENTER] key will display the selection for lines c and d. Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting ‘No’ will discard the entered values and return to the measurement display screen, selecting ‘Yes’ will save changes made. 8.3.2 Temperature source Press the [ENTER] key to select this Menu.
Transmitter pH M300 V20 38 Selecting No means the Temperature configured under 8.3.2 will be used for the Calibration. Selection Yes gives the user the Option to enter a fixed Calibration Temperature. 8.3.4 Set averaging Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each channel can now be selected.
Transmitter pH M300 8.4 V20 39 Analog outputs (PATH: Menu/Configure/Analog Outputs) Press the [ENTER key to select this Menu, which lets you configure the 2 Analog Outputs. The analog output number ( default is 1), the channel (a, b, c, d or blank (none) ) and a value (off, 3.6 mA or 22.0 mA. default is off) are configured. If a mA value is, the value of the analog output will go to the set value if any of the alarm conditions occur The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic.
Transmitter pH M300 8.5 V20 40 Set points (PATH: Menu/Configure/Set Points) Press the [ENTER] key to select this Menu. Up to 4 Set Points can be configured on any of the channels (a thru d). Possible types are Off, High, Low, Outside and Between. An ‘Outside’ Setpoint will cause an alarm condition whenever the measurement goes above its high limit or below its low limit. A ‘Between’ Setpoint will cause an alarm condition to occur whenever the measurement is between its high and low limits.
Transmitter pH M300 8.6 V20 41 Alarm/Clean (PATH: Menu/Configure/Alarm/Clean) This Menu allows the configuration of Alarm and Clean functionality. 8.6.1 Alarm Select the Relay to be used for the Alarm. Select one of the following events to be alarmed: 1. Power failure 2. Software failure 3. Rg Diagnostics 4. Rr Diagnostics. If any of these are set to yes then the alarm will come on and an alarm message will be recorded if: 1. there is a power failure or power cycling 2.
Transmitter pH M300 8.6.2 V20 42 Clean Configure the Relay to be used for the cleaning cycle. The Cleaning Interval can be set from 0.000 to 999.9 hours. Setting it to 0 turns the clean cycle is off. The cleaning time can be 0 to 9999 seconds and must be smaller than the Cleaning Interval. Select the desired Relay state: Normal or Inverted. Pressing the [ENTER] key again will bring up the Save Changes dialog.
Transmitter pH M300 8.7.2 V20 43 Resolution This menu allows the setting of the resolution of each displayed values. Possible settings are 1, 0.1, 0.01, 0.001 or Auto. Pressing the [ENTER] key again will bring up the Save Changes dialog. 8.7.3 Backlight This Menu allows the setting of the back light options of the display. Possible settings are On, On 50% or Auto Off 50%. If Auto Off 50% is selected then the backlight will go to 50% after 4 minutes with no keypad activity.
Transmitter pH M300 8.7.4 V20 44 Name This menu allows for the configuration of a text which is displayed on Lines 3 and 4 of the Display. The Name can be alphanumeric and is displayed in the last 9 characters of Line 3 and Line 4 during normal measurement mode. The default is nothing (blank). 8.8 Hold outputs (PATH: Menu/Configure/Hold Outputs) This Menu allows for the configuration of the Hold status.
Transmitter pH M300 9 V20 45 System (PATH: Menu/System) While in Measurement mode press the key. Press the or key to navigate to the System – Menu. After selecting the System – Menu enter the System security code and then press the [ENTER] key to confirm the code. 9.1 Set language (PATH: Menu/System/Set Language) This Menu allows the configuration of the Display language. Following selections are possible: English, French, German, Italian and Spanish.
Transmitter pH M300 9.3 V20 46 Passwords (PATH: Menu/System/Passwords) This Menu allows for the configuration of Operator and Administrator Passwords, as well as setting up a List of allowed Menus for the Operator. The Administrator has rights to access all Menus. The default passwords are 00000 for Administrator and Operator. The Passwords Menu is protected: Enter the Administrator Password to enter the Menu. 9.3.1 Changing passwords See 9.3 on how to enter the Passwords Menu.
Transmitter pH M300 V20 47 The Lockout – Menu is protected: Enter the Administrator Password and select YES to enable or NO to disable the Lockout functionality. Pressing the [ENTER] key after the selection will bring up the Save Changes dialog. 9.5 Reset (PATH: Menu/System/Reset) This Menu allows for following options: Reset System, Reset Meter Cal, Reset Analog Cal. 9.5.1 Reset system This Menu allows the reset of the meter to the factory default settings (see Default tables).
Transmitter pH M300 9.5.3 V20 48 Reset analog calibration This Menu allows the reset of the Analog Output calibration factors to the last factory calibration values. Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will discard the entered value, selecting Yes will make the entered value the current one.
Transmitter pH M300 10 V20 49 PID setup (PATH: Menu/PID Setup) PID Setup Tune Parameters Mode PID Display PID control is proportional, integral and derivative control action that can provide smooth regulation of a process. Before configuring the transmitter, the following process characteristics must be identified.
Transmitter pH M300 V20 50 To counteract this, the instrument allows for adjustable non-linear control with settings of a deadband around the setpoint, corner points further out and proportional limits at the ends of control as shown in the figure below. Determine the appropriate settings for each of these control parameters based on the shape of the process titration curve.
Transmitter pH M300 10.3 V20 51 Mode (Path MENU/PID Setup/Mode) This menu contains the selection of control modes using relays or analog outputs. Press [ENTER]. Relay(s) or analog output(s) for PID control action as well as details of their operation can be assigned. Based on the control device being used, select one of the following three paragraphs for solenoid valve, pulse input metering pump or analog control.
Transmitter pH M300 10.4 V20 52 Tune parameters (Path MENU/PID Setup/Tune Parameters) This menu assigns control to a measurement and sets the setpoint, tuning parameters and non-linear functions of the controller through a series of screens. 10.4.1 PID assignment & tuning Assign the measurement, a, b, c, or d to be controlled after “PID on_”. Set the Gain (unitless), Integral or Reset time Tr (minutes) and Rate or Derivative time Td (minutes) needed for control. Press [ENTER].
Transmitter pH M300 10.5 V20 53 PID display setup (Path MENU/PID Setup/PID Display This display enables display of PID control status including % output, reagent and Auto/Manual, in the normal measurement mode. (In addition, for the display to be enabled, a measurement must be assigned under Tune Parameters and a relay or analog output must be assigned under Mode.) If Yes is selected the status (Man or Auto) and control output (% acid or % base) will be displayed on the bottom line.
Transmitter pH M300 11 V20 54 Service (PATH: Menu/Service) While in Measurement mode press the key. Press the or key to navigate to the Service – Menu. After selecting the Service – Menu enter the Service security code and then press the [ENTER] key to confirm the code. 11.
Transmitter pH M300 V20 55 11.1.2 Digital input The digital Input menu shows the state of the digital input. Press [ENTER] to exit from this display. 11.1.3 Display All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After 15 seconds the transmitter will return to the normal measuring mode or press [ENTER] to exit sooner. 11.1.4 Keypad For the keypad diagnostics the display will indicate which key is pressed.
Transmitter pH M300 V20 56 11.1.6 Set relay The Set Relays diagnostic menu allows for the manual activation/deactivation of each Relay. 11.1.7 Read relays The Read Relays diagnostic menu shows the state of each Relay. Press [ENTER] to exit from this display. 11.1.8 Set analog outputs This menu enables the user to set all analog outputs to any mA value within the 022 mA range.
Transmitter pH M300 V20 57 11.1.9 Read analog outputs This menu shows the mA value of the analog Outputs. Press [ENTER] to exit from this display. 11.2 Calibrate (PATH: Menu/Service/Calibrate) This menu has the options to calibrate the transmitter and the analog outputs and also allows the unlocking of calibration functionality. 11.2.1 Calibrate meter Channel A calibration can be selected as Voltage, Temperature, Rg Diagnostic and Rr Diagnostic.
Transmitter pH M300 V20 58 11.2.1.1 Voltage Enter the value for Point 1 of the calibration. Enter the value for Point 2 of the calibration. Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones. 11.2.1.2 Temperature Enter the value for Point 1 of the calibration, according to the attached Resistor.
Transmitter pH M300 V20 59 11.2.1.3 Rg diagnostic Enter the value for Point 1 of the calibration according to the resistor connected across the pH glass electrode measuring input. Enter the value for Point 2 of the calibration according to the resistor connected across the pH glass electrode measuring input. Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Selecting No will discard the entered values, selecting Yes will make the entered values the current ones. 11.2.
Transmitter pH M300 V20 60 As the five digit number is increased the output current increases and as the number is decreased the output current decreases. Thus coarse changes in the output current can be made by changing the thousands or hundreds digits and fine changes can be made by changing the tens or ones digits. Pressing the [ENTER] key after entering both values will bring up a confirmation screen.
Transmitter pH M300 12 V20 61 Info (PATH: Info) Pressing the Calibration Data 12.1 key will display the Info Menu with the options Messages and Messages (PATH: Info/Messages) The most recent message is displayed. The up and down arrow keys allow scrolling through the last four messages that have occurred. Clear Messages clears all the messages. Messages are added to the message list when the condition that generates the message first occurs.
Transmitter pH M300 12.3 V20 62 Model/software revision Selecting Model/Software Revision will display the installed Firmware revision. The displayed information is important for any Service call. Press Enter to return to the normal measurement mode.
Transmitter pH M300 13 V20 63 Maintenance The device contains no user repairable components. To remove dust, dirt and spots, the external surfaces of the transmitter may be wiped with a damp, lint-free cloth. 13.1 For technical support For technical support and product information contact your local Mettler-Toledo Sales Office or representative 13.2 Front panel cleaning Clean the front panel with a damp soft cloth (water only, no solvents). Gently wipe the surface and dry with a soft cloth.
Transmitter pH M300 14 V20 64 Trouble shooting If the equipment is used in a manner specified by these instructions, the protection provided by the equipment may be impaired. Review the table below for possible causes of common problems: Problem Possible cause Display is blank. - Incorrect measurement readings. Measurement readings not stable. - Cannot change menu settings. - No power to M300 Blown fuse. LCD display contrast set incorrectly. Hardware failure. Sensor improperly installed.
Transmitter pH M300 14.1 V20 65 Warning- and alarmlist Warnings Description Warning pH slope > 101% Slope too big Warning pH Slope < 95% Slope too small Warning pH Zero >7.5 pH Zero offset too big Warning pH Zero < 6.5pH Zero offset too small Warning pHGls change < 0.3 Glass electrode drift too small Warning pHGls change > 3 Glass electrode drift too big Warning pHRef change < 0.
Transmitter pH M300 15 V20 66 Accessories and spare parts Please contact your local Mettler-Toledo sales office or representative for details on available accessories and spare parts. 16 Specifications 16.1 General specifications pH range pH Resolution pH Relative accuracy mV range mV Resolution mV Relative accuracy Temperature measuring range Temperature resolution Temperature measurement error Temperature input Temperature compensation Calibration 16.2 -1.00 to 15.00 pH 0.01 pH ± 0.
Transmitter pH M300 16.3 90 x 90 x 140 mm (1/4DIN model) 102 x 102 mm 125 mm (excludes plug-in connectors) 0.6kg ABS/polycarbonate IP 65 (front) / IP 20 (housing) Mechanical specifications for 1/2DIN version Dimensions (housing - L x H x W)* Front bezel - H x W Max. D - panel mounted Weight Material Insulation / rating * H=Height, W=Width, D=Depth 16.5 67 Mechanical specifications for 1/4DIN version Dimensions (housing - H x W x D)* Front bezel – (H x W) Max.
Transmitter pH M300 17 V20 68 Default tables Alarm (except flow) Clean Language Passwords All Relays (unless otherwise specified) Lockout Channel A Cal constants pH Buffer Analog Out All analog out pH Temperature Set point 1 Relay 3 Set point 2 Relay 4 Resolution Relay diagnostics SW HW delay hysteresis state relay hold mode interval clean time state delay hysteresis administrator operator delay hysteresis state hold mode (on/off) measurement a measurement b measurement c measurement d pH tempe
Transmitter pH M300 18 V20 69 Warranty METTLER TOLEDO warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. METTLER TOLEDO's Customer Service Dept. will determine if the product problem is due to deviations or customer abuse.
Transmitter pH M300 20 V20 70 Buffer tables M300 transmitters have the ability to do automatic pH buffer recognition. The following tables show different standard buffers that are automatically recognized. 20.1 Mettler-9 Temp (°C) pH of buffer solutions 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 2.03 2.02 2.01 2.00 2.00 2.00 1.99 1.99 1.98 1.98 1.98 1.98 1.98 1.99 1.99 2.00 2.00 2.00 2.00 2.00 4.01 4.01 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.04 4.06 4.08 4.10 4.13 4.16 4.19 4.22 4.
Transmitter pH M300 20.2 V20 Mettler-10 Temp (°C) pH of buffer solutions 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 2.03 2.02 2.01 2.00 2.00 2.00 1.99 1.99 1.98 1.98 1.98 1.98 1.98 1.99 1.99 2.00 2.00 2.00 2.00 2.00 20.3 71 4.01 4.01 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.04 4.06 4.08 4.10 4.13 4.16 4.19 4.22 4.26 4.30 4.35 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.04 7.06 7.09 7.12 10.32 10.25 10.18 10.12 10.06 10.01 9.97 9.93 9.89 9.86 9.
Transmitter pH M300 20.4 V20 72 NIST standard buffers (DIN 19266: 2000-01) Temp (°C) pH of buffer solutions 0 5 10 15 20 25 30 35 40 45 50 55 60 70 80 90 95 1.668 1.670 1.672 1.676 1.680 1.685 1.694 1.697 1.704 1.712 1.715 1.723 1.743 1.766 1.792 1.806 4.004 4.001 4.001 4.003 4.008 4.015 4.028 4.036 4.049 4.064 4.075 4.091 4.126 4.164 4.205 4.227 6.950 6.922 6.900 6.880 6.865 6.853 6.841 6.837 6.834 6.833 6.834 6.836 6.845 6.859 6.877 6.886 9.392 9.331 9.277 9.228 9.184 9.144 9.095 9.076 9.046 9.
Transmitter pH M300 20.5 V20 73 Hach buffers Buffer values up to 60 °C as specified by Bergmann & Beving Process AB. Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 *Values complemented 20.6 pH of buffer solutions 4.00 4.00 4.00 4.00 4.00 4.01 4.01 4.02 4.03 4.05 4.06 4.07 4.09 4.09* 4.09* 4.09* 4.09* 4.09* 4.09* 4.09* 7.14 7.10 7.04 7.04 7.02 7.00 6.99 6.98 6.98 6.98 6.98 6.98 6.99 6.99* 6.99* 6.99* 6.99* 6.99* 6.99* 6.99* 10.30 10.23 10.11 10.11 10.05 10.00 9.96 9.92 9.88 9.
Transmitter pH M300 20.7 V20 Merck Titrisols, Reidel Fixanals Temp (°C) pH of buffer solutions 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 2.01 2.01 2.01 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.01 2.01 2.01 2.01 2.01 2.01 20.8 74 4.05 4.05 4.02 4.01 4.00 4.01 4.01 4.01 4.01 4.01 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 4.00 7.13 7.07 7.05 7.02 7.00 6.98 6.98 6.96 6.95 6.95 6.95 6.95 6.96 6.96 6.96 6.96 6.97 6.98 7.00 7.02 9.24 9.16 9.11 9.05 9.00 8.95 8.91 8.
Transmitter pH M300 V20 75 Notes: © 02/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter pH M300 52 121 304
TD_InPro8600_Trb8300D_E_52800992 14.12.2006 11:20 Uhr Seite 16 METTLER TOLEDO Market Organizations Sales and Service: Australia Mettler-Toledo Ltd. 220 Turner Street Port Melbourne AUS - 3207 Melbourne / VIC Phone +61 31300 659 761 Fax +61 3 9645 3935 e-mail mtausprocess@mt.com Austria Mettler-Toledo Ges.m.b.H. Südrandstraße 17 A - 1230 Wien Phone +43 1 604 19 80 Fax +43 1 604 28 80 e-mail infoprocess.mtat@mt.com Brazil Mettler-Toledo Ind. e Com. Ltda.