Specifications

Transmitter pH M300 V20 40
8.5 Set points
(PATH: Menu/Configure/Set Points)
Press the [ENTER] key to select this Menu.
Up to 4 Set Points can be configured on any of the channels (a thru d). Possible
types are Off, High, Low, Outside and Between.
An ‘Outside’ Setpoint will cause an alarm condition whenever the measurement
goes above its high limit or below its low limit. A ‘Between’ Setpoint will cause an
alarm condition to occur whenever the measurement is between its high and low
limits.
Enter the desired value(s) for the Set Point
This screen provides the option to configure a setpoint to be active on an over range
condition. Select the setpoint (1 thru 4) and ‘Yes’ or ‘No’. Select the desired relay
that will activate when the setpoint alarm condition is reached.
Set the Relay delay and the Hysteresis of the configured Relay.
Set the type of Relay contact, either normal or inverted.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting
No will discard the entered values, selecting Yes will make the entered values the
current ones.