Operating instructions

¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
Operating hours
10y 321d 23h
Total weight loaded
1485.1345 t
Number of weights
999999999999
Number of key press
999999999999
Motor movements
9999999999
Backlight runnig hours
10y 321d 23h
Next service due date
01.01.2010
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦ ------------------------ ¦
16.6 Calibration History
The "Calibration History" function allows you to view and print information on the last 30 (thirty) balance adjust
ment. Adjustments made by a service technician and normal user are counted together.
a) Press « » to start "CAL.HIST".
b) Press « » key to scroll forward through the displayed list of Adjustments history information.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOSE".
d) Press « » for printout.
Sample Information Displayed:
Note Display
S = External adjusted service 06:03:09 S
I = Internal adjusted 05:03:09 I
·
·
·
F = FACT 04:03:09 F
I = Internal adjusted 03:03:09 I
E = External adjusted user 02:03:09 E
I = Internal adjusted 05:03:09 I
Sample Printout:
¦ ------ Calibration ------ ¦
¦ 05.Mar. 2009 11:34 ¦
¦
¦ METTLER TOLEDO
¦
¦
¦
¦
¦
Balance Type
SNR:
MS204S
1234567890
¦
¦
¦
¦ --------------------------- ¦
MS-S / MS-L Models62