Operating instructions

AT/MT/UMT Balances METTLER TOLEDO
20
Function: Triggers the automatic calibration or calibration test, changes
the calibration mode, returns calibration status.
Format: CA[ n|U|S|T|?]
CA
Calibration
Description:
n
is either 0 or 1.
CA starts the automatic internal calibration and linearization.
CA 0 sets the mode CALIN “Auto off” (the internal calibration is
launched by the user or through the CA command).
CA 1 sets the mode CALIN “Auto on” (fully automatic calibration).
CA U sets the mode CAL “User” (calibration with external reference
weight).
CA ? sends the currently set mode as follows:
CA=0 mode “Auto off”
CA=1 mode “Auto on”
CA=U mode “User”
A calibration operation in progress can be aborted with the cancel
command “.” .