Operating instructions
AT/MT/UMT Balances METTLER TOLEDO
22
Function: Sets all balance parameters, except those for the serial
interface, to the factory setting.
Format: CFD
CFD
Configuration Defaults
Description: CFD essentially causes resetting of the balance to the standard
configuration. The following parameters are affected:
• vibration adapter (wave symbol)
• weighing process adapter (drop symbol)
• stability detector (ASD)
• AutoZero
• automatic draft shield operation
• calibration mode
• automatic turning-off of the status displays
• sound generation (on or off)
• balance resolution (readout increment)
• weight units 1 and 2
• “nationality” of weight unit tael
The parameters for the serial interface remain unchanged in order to
avoid disrupting communication with the peripheral device.