Operating instructions

METTLER TOLEDO AT/MT/UMT Balances
25
Description: After CFP the balance sends the contents of the configuration register
as normal text. Note that this information does not necessarily reflect
the
current
configuration since that could have been changed
temporarily by interface commands. The configuration supplied is
that which is permanently stored in the balance, and which is valid
immediately after the balance has been switched on. The actual
current settings may be recalled by the appropriate commands (e.g.
MS ?).
A typical configuration is transmitted as follows:
CFP
Configuration Print
Function: Prompts the balance to send the contents of the
configuration register.
Format: CFP
S T O R E D S E T T I N G S
SCALE:
Stability Detector (ASD): -3-
Display Resolution (dC) : step by 1
Display Resolution (dF) : step by 1
AutoZero : on
Calibration Method : internal auto
UNIT:
Unit 1 : g
Unit 2 : %
Display Symbols : auto-off
Sound : on
INTERFACE:
Weight Send Method : stable only (S)
Baudrate : 2400
Parity : even
Handshake Method : Pause after line
End-of-line : CRLF
——
(coarse range)
(fine range)