Operating instructions

AT/MT/UMT Balances METTLER TOLEDO
62
Description: Format 1:
stat
is a natural number in the range 0 to 255.
With this command the eight digital outputs of the GM54 Output
Module accessory can be switched on and off individually. Each of
the eight outputs is assigned a value as follows:
digital output 0: 1
digital output 1: 2
digital output 2: 4
digital output 3: 8
digital output 4: 16
digital output 5: 32
digital output 6: 64
digital output 7: 128
stat
is specified as the sum of the values of those switches that
should be closed. Example: With W 5 the digital outputs 0 and 2 are
closed, all others are opened.
W without parameter opens all eight switches.
A break in the interface line also opens all switches.
Error: EL if
stat
is not in the range 0 to 255.
W
Write Output
Function: Controls the GM54 Output Module or the LV10 Automatic
Feeder.
Format 1: W[
stat
]
Format 2: W
stat time stat
[
time stat
]…