Operating instructions

AT/MT/UMT Balances METTLER TOLEDO
70
Interface Configuration
All parameters that control the behavior of the interface can be selected in the fourth sector
of the configuration register. The four send modes described below define the behavior of
the balance after switching on or after an interface break. The send mode can be overwritten
at any time by send commands.
Send Stable (“S.Stb”)
After the key “Print” or the external transfer key has been pressed, the next stable result is
transmitted. A brief beep sounds at the same time. If the fully automatic door function
(“doorAuto”) has been selected in the menu, the draft shield first closes automatically and
then reopens after transmission.
To distinguish data strings initiated this way from those resulting from an “S” command,
the first character is a space instead of an “S”.
Send All (“S.ALL”)
After the “Print” or the external transfer key has been pressed, a result is transmitted
immediately without closure of the draft shield. Stable values are identified by a beep.
To distinguish data strings initiated this way from those resulting from an “SI” command,
the first character is a space instead of an “S”.
Send Continuous (“S.Cont”)
The balance continuously transmits approximately two and a half results per second . The
“Print” key and the external transfer key have the same effect as in the “Send Stable” mode.
The data format is not distinguished from that following an “SIR” command.