Operating instructions

Basic information on data interchange with the balance
The balance receives commands from the system and acknowledges the command with an appropriate
response.
Command formats
Commands sent to the balance comprise one or more characters of the ASCII character set. Here, the following
must be noted:
Enter commands only in uppercase.
The possible parameters of the command must be separated from one another and from the command
name by a space (ASCII 32 dec., in this description represented as ).
The possible input for "text" is a sequence of characters of the 8-bit ASCII character set from 32 dec to 255
dec.
Each command must be closed by C
R
L
F
(ASCII 13 dec., 10 dec.). The characters C
R
L
F
, which can be
inputted using the Enter or Return key of most entry keypads, are not listed in this description, but it is
essential they be included for communication with the balance.
Example
S – Send stable weight value
Command S Get the current stable net weight value.
Response S S WeightValue Unit Current stable weight value in unit actually set
under unit 1.
S I Command not executable (balance is currently
executing another command, e.g. taring, or
timeout as stability was not reached).
S + Balance in overload range.
S - Balance in underload range.
Example
Command S Query a stable weight value.
Response S S 100.00 g The current stable weight value is 100.00 g.
The available MT-SICS commands are listed in the table (depending on the model). For further information
please refer to the Reference Manual "MT-SICS 11780711" downloadable from the Internet under
www.mt.com/sics-newclassic.
Description Description
@ Cancel (Reset) M46 Print interval
C0 Query/Set adjustment settings PW Piece counting: Query/Set piece weight
C1 Start adjustment according to current set
tings
PWR Power on/off (PWR 0 means switch off
completely, if balance is powered by bat
tery)
C2 Start adjustment with external weight S Send stable weight value
C3 Start adjustment with internal weight SI Send weight value immediately
D Display text sent to balance SIR Send weight value immediately and repeat
DAT Date query/set SIRU Send weight value with currently displayed
unit immediately and repeat
DW Display weight SIU Send weight value with currently displayed
unit immediately
I0 Commands implemented SM0 Dynamic weighing: cancel all SMx com
mands
I1 MT-SICS level and MT-SICS versions SM1 Dynamic weighing: Start immediately and
send the result
I2 Balance data SM2 Dynamic weighing: start after a minimum
load is exceeded and send result
87Interface Specification