To protect your METTLER TOLEDO product‘s future: METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come. Please send for full details about our attractive terms of service. Thank you. *P11780223* Subject to technical changes and to the availability of the accessories supplied with the instruments. © Mettler-Toledo GmbH 1999 11780223A Printed in Switzerland 9911/2.
1 2 3 1 MT5 Max 5,1 g d = 1 µg Pro FACT 4 5 6 7 8 METTLER TOLEDO Cancel 10/1 µg AUTOCALIN REMOTE STANDBY AZ G B N 2 T Configuration Print Menu 11 11 Select 2 Select 1 Set 16 On/Off Re-Zero Re-Zero Set 9 10 12 On/Off P r i n t M e n u 1 15 13 3 4 5 10 17 14 19 18 18 6 7 8 9 21 MT5 Max 5,1 g d = 1 µg 22 Pro FACT METTLER TOLEDO Cancel 10/1 µg AUTOCALIN REMOTE STANDBY AZ G B N T Configuration Print Menu 20 20 Select 1 Select 2 Set Re-Zero 23 On/Of
Operating instructions METTLER TOLEDO MT/UMT balances 1 Your balance is distinguished primarily, however, by its user friendliness: You attach the weighing cell to the evaluation unit, connect the balance to the power supply and after 60 minutes warm-up time (with an acclimatized balance) you can immediately start weighing. You thus obtain your first weighing result in an extremely short time. Your balance smooths the way 1.
2 1.2 Preparing your balance If you fold out the front fold-out, you have in front of you an immediate overview of all general illustrations at a glance. You need undertake the following operations only on first-time installation and each time you change the balance location. - No direct sunlight - No extreme temperature fluctuations - No extreme fluctuations in the atmospheric humidity - No excessive drafts (e.g.
Operating instructions METTLER TOLEDO MT/UMT balances Unpack your balance Connect the evaluation unit and the weighing cell - Plug the supplied connection cable into connection socket 7 of the evaluation unit (marked with “For connection to weighing cell only”) and plug the other end into connection socket 19 of the weighing cell (marked with “For connection to evaluation unit only”). It is immaterial which end of the cable you connect to the evaluation unit or the weighing cell.
4 Mount the weighing pan and the draft shield Level your balance - - - The MT balances are supplied with mounted weighing pan. You thus only need to mount the glass cover on draft shield 12 of the weighing cell. The UMT balances have a smaller weighing pan than the MT balances (to avoid cornerload errors). The weighing pan is already mounted, on delivery. You thus need only mount the glass cover on the draft shield 12 of the weighing cell.
Operating instructions METTLER TOLEDO MT/UMT balances STANDBY After connection, your balance is on standby. If your balance is already acclimatized (depending on the temperature difference between the old and new location, this takes 6 to 12 hours), it provides you with the first reliable results after 60 minutes warmup time. After approximately 2 hours you can count on very accurate weighing results. 1.
6 Your balance is fitted with a fully automatic door function, which ensures that the draft shield is always open or closed at the right time. Section 3.4 describes how you can switch off this automatic mechanism. On completion of the self-test, this function opens the draft shield completely automatically. ------ 0.000 mg Zeroing of the display (taring) is automatic.
Operating instructions METTLER TOLEDO MT/UMT balances - When the symbol of the stability detection (ASD) fades, the warning beep sounds; the triangle symbol also fades and the draft shield opens automatically. - Read off the weighing result. The result remains “frozen” on the display for 5 seconds. This can be recognized by the flashing circle above the weight unit. If the balance is attached to a printer, the weighing result is automatically printed out. 5.346 mg 5.
8 Some additional tips: - If you have made a mistake during weighing, there is no need to worry: Simply switch off your balance briefly and then switch it on again. You will later learn of other possibilities to bring your MT/UMT to weighing readiness again. - You can also effect the zeroing (taring) of the balance using the foot or hand switch (available as an accessory) attached to the balance via connection 3. Operation with the foot switch leaves both your hands free for loading the balance.
Operating instructions METTLER TOLEDO MT/UMT balances 2 Your balance offers high-level operating convenience MT5 Max 5,1 g d = 1 µg The semiautomatic operation can be effected either with the or key. In this case, the draft shield is opened and closed by the motor either to the right or left. Pro FACT Cancel 10/1 µg Configuration Menu Select 1 Set 2.
10 The fully automatic door function primarily facilitates work with the single or twin foot switch. Section 3.4 shows you how to switch off the fully automatic door function. The draft shield of your balance has a “Learn function”. If the fully automatic door function is in operation, the draft shield opens automatically when the balance is zeroed and when the weighing result is printed out on the side you desire.
Operating instructions METTLER TOLEDO MT/UMT balances 24.083 98.215 2.2 The DeltaTrac® - an informative display 2.3 How you can weigh in the coarse and fine range The DeltaTrac® 22 is a supplement to the digital display 23. It provides you with a pictorial representation of the weighing range in use and that still remaining. The dynamics of a weighing process can be followed on the DeltaTrac® extremely well.
12 You have the following selectable ranges available: Balance Fine range Coarse range MT5 1 µg 0.01 mg 10 / 1 µg UMT2 0.1 µg 1 µg 1 / 0.1 µg UMT5 0.1 µg 1 µg 1 / 0.1 µg - ...the balance operates in the coarse range. - By pressing the «10/1 µg» (or «1/0.1 µg») key again, you can return to the fine range. 12.644 mg Key With the «10/1 µg» key of the MT balance or the «1/0.
Operating instructions METTLER TOLEDO MT/UMT balances 3 Overview of the menu functions: - reference weight in % - reference weight in pieces - adjustment - weighing process adapter - vibration adapter - fully automatic door function You will find a detailed overview of the menu on pages 78 and 79. Your balance is an adaptable balance For the optimization of your weighing results and adaptation to your weighing needs, your balance offers you additional setting options.
14 In addition, you can also perform several operational steps in the menu for special applications (percent weighing and piece counting). Section 5 of these operating instructions provides you with more information. MT5 Max 5,1 g d = 1 µg Cancel 10/1 µg Configuration Menu Select 1 Set Pro FACT We encourage you to experiment with the menu to familiarize yourself with it.
Operating instructions METTLER TOLEDO MT/UMT balances MT5 Max 5,1 g d = 1 µg Pro FACT By repeated pressing of the «Menu» key, all available menu options appear in the display in succession: «Cancel» key Cancel 10/1 µg Configuration Menu Select 1 Set - Exit the menu and return to the weighing mode without any changes being made. If you have made a wrong selection, you can leave the menu at any time by pressing the «Cancel» key.
16 A complete overview of the menu can be found on pages 78 and 79. - Some additional tips: - Not all menu options are constantly or exactly displayed in the form shown; some of them, e.g. the setting of reference weights (“Set 100%”) depend on additional settings in the configuration register, which we will introduce you to in Section 4 of these operating instructions. - If you are in the menu and press no key for about 1 minute, the balance returns automatically to the weighing mode.
Operating instructions METTLER TOLEDO MT/UMT balances This is the universal setting suitable in principle for all types of weighing. This setting is preset in the factory. 3.2 How to adapt your balance to different types of weighing Your balance can be matched to different types of weighing (fine dispensing, absolute weighing, etc.) very simply. For this purpose, it is equipped with the so-called weighing process adapter, which can be set in the menu.
18 In this position the drop symbol is invisible, the weighing process adapter is switched off. This setting is suitable only for special applications in which not only the final result but also the dynamics of the weighing process are of interest, for example the recording of weight changes due to evaporation against time. MT5 Max 5,1 g d = 1 µg Cancel 10/1 µg Configuration Menu Select 1 Set Pro FACT - 3.
Operating instructions METTLER TOLEDO MT/UMT balances MT5 Max 5,1 g d = 1 µg Cancel 10/1 µg Pro FACT - Select the appropriate setting with the «Select 1» key: MT5 Max 5,1 g d = 1 µg Cancel 10/1 µg Configuration Configuration Menu Menu Select 1 Select 1 Set Set Pro FACT - When you have selected the appropriate setting, press the «Set» key. The chosen setting then immediately becomes active, and the balance returns to the weighing mode.
20 When the twin foot switch is used, the automatic operation leaves both hands free for weighing in. 3.4 How to switch off the fully automatic door function You are already familiar with the convenient automatic door function of your balance from your first weighing. When the automatic door operation is switched on, the draft shield closes automatically after the «Re-Zero» key (zeroing/ taring) or the «Print» key (printout of the weighing result, see Section 5.6) has been pressed.
Operating instructions METTLER TOLEDO MT/UMT balances MT5 Max 5,1 g d = 1 µg Pro FACT - Cancel 10/1 µg Configuration Menu Using the «Select 1» key, you can now switch the automatic door operation off and on again: 4 Additional possibilities of your balance Select 1 In addition to the menu, your MT/UMT balance offers you further, very specific adaptation possibilities. You perform these adaptations in the configuration register, which we will introduce you to in the following section.
22 Unit Int -FACE - Weighing units weighing unit 1 weighing unit 2 status indicator beeper - Interface data transmission mode data transmission rate parity transmission protocol end-of-line mode You will find further details on the adjustment parameters on pages 78 and 79. 4.2 How is the configuration register used? Hold the short-form operating instructions in your hand. You will then have the complete configuration register in front of you at a glance.
Operating instructions METTLER TOLEDO MT/UMT balances MT5 Max 5,1 g d = 1 µg By pressing and holding the «Configuration» key, you call up the configuration register. You select the different - sectors («Configuration» key) - parameters («Select 1» key) - settings («Select 2» key) by simply pressing the appropriate key. In contrast to the menu, you also need the «Select 2» key for operation of the configuration register.
24 MT5 Max 5,1 g d = 1 µg Pro FACT A further tip: «Cancel» key Cancel 10/1 µg Configuration Menu Select 1 Set 2400 bd Exit the configuration register and return to the weighing mode without any changes being stored. If you have made a wrong selection, you can quit the configuration register at any time by pressing the «Cancel» key. You are then again in the weighing mode without any setting having been altered.
Operating instructions METTLER TOLEDO MT/UMT balances 4.3 In the first sector you return to the factory settings - Select the first (and only) parameter («Select 1» key). - Activate the resetting by pressing the «Set» key, and the scale returns automatically to the weighing mode. You again work with the factory settings. rESEt But here all specific settings, even those in the menu, will be replaced by the factory settings! 0.
26 SCALE 4.4 The second sector offers you numerous adaptation possibilities The symbol fades and the weighing result is released only when the result is within the limits selected by you for the stability. - You have a choice of eight settings. In the setting Asd–1– the display must be within a preset tolerance limit for one second for the result to be released as stable. In the setting ASd–7– the result must be within a preset tolerance limit for several seconds before being released.
Operating instructions METTLER TOLEDO MT/UMT balances Stage Weighing speed Repeatability d In this parameter you define the readout increments (digits) for the last decimal place. Readout increments of 1, 2, 5 and 10 are available, and these are symbolized by the following setting options: off very fast 1 2 3* 4 5 6 7 fast Readability (selection of the readout increments) good very good MT5 UMT2 UMT5 0.01 mg* 0.02 mg 0.05 mg 0.
28 dF Readability (selection of the readout increments) in the fine range Autozero (AZ) AZ In this parameter you switch the automatic zeroing (autozero) on or off. When switched on, the zero is automatically corrected for drift or contamination of the weighing pan. Here you define the readout increments (digits) for the last decimal place for weighings in the fine range.
Operating instructions METTLER TOLEDO MT/UMT balances CALIN proFACT professional fully automatic adjustment and linearization technology Adjustment and linearization with proFACT In this parameter you define the adjustment mode, i.e. the way in which your balance should be adjusted and linearized. The following possibilities are available: CALIN Auto on - The MT/UMT balance adjusts and linearize itself fully automatically by loading two internal adjustment weights.
30 In this mode you can also initiate the fully automatic adjustment and linearization at any time with a keystroke. This is practical, for instance, when your balance reports during a weighing that it wishes to adjust itself. Section 5.4 shows you how to proceed in such a case. CALIN Auto oFF - CAL - USEr The fully automatic self-adjustment with linearization adjustment (proFACT) described above is switched off. You can trigger the adjustment with the internal adjustment weights at a keystroke.
Operating instructions METTLER TOLEDO MT/UMT balances The number of decimal places in each unit is dependent on the balance model. 4.5 In the third sector you select the weight units and additional settings In this sector you can select the following parameters and their settings: By pressing the “ key, you toggle between weight unit 1 (Unit 1) and weight unit 2 (Unit 2 ). Section 5.1 provides you with further details regarding this switching.
32 mo PCS Stk % Status indicators momme pieces pieces (German) percent The status indicators 21, such as the symbols for the vibration adapter and the weighing process adapter, provide you with an overview of the selected settings. These graphic indicators can be switched on or off. * = factory setting The number of decimal places in each unit depends on the balance model. By pressing the “ key, you toggle between weight unit 1 (Unit 1) and weight unit 2 (Unit 2 ).
Operating instructions METTLER TOLEDO MT/UMT balances Beeper setting bEEP 4.6 The fourth sector presents a communication-friendly balance The audio acknowledgement beep supports the operation of the balance and the display audibly. You can switch it off or on. Int-FACE bEEP on bEEP oFF audio acknowledgement beep switched on * Select the fourth sector in the configuration register (interface).
34 S. Data transmission mode Important note: The following setting options are available: If you have switched off the automatic stability detection in sector 2 (ASD oFF ), all weight values are regarded as being stable. After the «Print» key has been pressed in this case, the following values are transmitted: S. Stb Send stable: The first stable value is transferred after initiation of the print command after release by the automatic stability detection. * In setting S.
Operating instructions METTLER TOLEDO MT/UMT balances Parity Data transmission rate (baud rate) bd 2400 bd PAr The data transmission rate (baud rate) determines the speed of the transmission via the serial interface. The unit is the baud (1 baud (bd) = 1 bit/second). The following transmission rates are available: - 150 db 300 db 600 bd 1200 bd 2400 bd * 4800 bd 9600 bd With the aid of the parity test, simple bit errors in the data transmission can be recognized by a parity check.
36 HS These settings allow matching of the data transmission to various serial receivers. The following settings are available: HS PAUSE HS CL - Pause of 1 second (for METTLER TOLEDO GA42 Printer)* METTLER TOLEDO transmission protocol. In this operating mode, the end-of-line mode is automatically set to “crLF ” and can not be changed.
Operating instructions METTLER TOLEDO MT/UMT balances Important note: Proceed as follows for the printout: If HS CL is set in the transmission protocol, the end-of-line mode is set automatically to EOL crLF and can not be changed. - Select the configuration register. - Press the «Print» key, and the current configuration will be printed out. METTLER TOLEDO Print Select 2 On/Off 4.7 Print out the current configuration STD 10.1.11 TYPE : MT 5 1.00.
38 4.8 How to protect your settings against inadvertent change As you know, with one press of the «Set» key your individual settings in the configuration register are stored. These are available at any time (even if the balance has since been disconnected from the power supply). You are not, however, protected against inadvertent alterations, for instance by other balance users. - Grasp the bracket of program cassette 8 and withdraw this carefully from the plug-in slot 6.
Operating instructions METTLER TOLEDO MT/UMT balances - Slide the program cassette 8 into the plug-in slot 6 up to the stop and replace the cover. Plug the cable of the AC adapter into the receptacle. 5 Your balance is also prepared in the best possible way for special cases. Applications and functions built in as standard expand the possibilities of your balance and facilitate the operation. The following section familiarizes you with these applications and functions.
40 68.432 µ METTLER TOLEDO You have, for example, selected “mg” as weight unit 1 and “ct” as weight unit 2 in the configuration register. You can switch between these two units at a keystroke: 5.2 Your balance also works with percentages and shows plus/minus deviations The balance is operating with weight unit 1. This built-in application allows you to weigh in to a preset value (100%) and to determine deviations from a target value. - Press key “ briefly and ... - ...
Operating instructions METTLER TOLEDO MT/UMT balances Note: This request appears only if weight unit 2 is set to percent (%). SEt 100 % - - Open the draft shield and load the reference weight. -2.5% 100% +2.5% - ------ Press the «Set» key and the draft shield closes automatically if the automatic door function is switched on. If not, close the draft shield manually or by pressing one of the two keys or and then press key «Set».
42 67.40 % coarse indicator Add the weighing sample at a steady rate to the container until the coarse indicator on the DeltaTrac® is pointing vertically downward. This shows that the approximate target weight has now been reached. 5.3 Do you wish to count pieces? First your balance needs to know that you wish to perform piece counting. You make this setting in the third sector of the configuration register (Unit ), where you set weight unit 2 (Unit 2 ) to “PCS ” (English) or “Stk ” (German).
Operating instructions METTLER TOLEDO MT/UMT balances - If you wish to count out of a container, load the full container. MT5 Max 5,1 g d = 1 µg - Pro FACT Cancel 10/1 µg Configuration Menu Select 1 Select the desired reference piece number with the «Select 1» key. The following piece numbers are available: Set 10 SEt 10 MT5 Pro FACT Max 5,1 g d = 1 µg METTLER TOLEDO - Set the scale to zero (taring).
44 - If the automatic door function is switched on, the draft shield opens automatically on completion of the weighing-in process. If not, open it yourself. If you wish to count out of a container, remove the selected number of pieces from the container. Depending on how you have weighed in the reference weight, you can now: - Press the «Set» key, and the draft shield closes automatically if the automatic door function is switched on.
Operating instructions METTLER TOLEDO MT/UMT balances You select the adjustment mode in sector 2 (SCALE ) of the configuration register (see Section 4.4). In addition to the fully automatic self-adjustment, you have two further adjustment possibilities available: 5.4 How to trigger the adjustment manually AUTOCALIN 0.367 In the factory the balance is preset to proFACT. The MT/UMT balance adjusts and linearizes itself as soon as a change in the ambient conditions makes this necessary.
46 Automatic adjustment with the internal adjustment weights at a keystroke You have this adjustment mode available if you have set the parameter “adjustment and linearization” in sector 2 of the configuration register (basic settings) either to CALIN Auto on or CALIN Auto oFF. AUTOCALIN - CAL Int - CAL End Repeatedly press the «Menu» key briefly until the adjacent display appears. Completion of the adjustment and linearization process is displayed briefly,... ...
Operating instructions METTLER TOLEDO MT/UMT balances Manual adjustment with an external adjustment weight CALIN 5.000000 You wish to use your own external adjustment weight for the adjustment and have set the parameter proFACT (adjustment and linearization) to CAL USEr in sector 2 of the configuration register (basic settings). In this mode, the automatic adjustment with the built-in adjustment weights is switched off.
48 Note: This display does not appear if you have set the parameter “adjustment and linearization” to CAL USEr in the configuration register (adjustment with your own external adjustment weight). Please also see the note at the end of this section. 5.5 How to check whether the balance is still adjusted correctly In the factory, the balance is set to proFACT, professional fully automatic adjustment and linearization technology. A check of the adjustment is not necessary in this mode.
Operating instructions METTLER TOLEDO MT/UMT balances 5.000001 The internal weights are now loaded, without the balance being adjusted, and the current, model-dependent adjustment value displayed, interrupted by the display tESt. 5.000000 5.6 How to print out the weighing result You can output the weighing result via the interface connection 4 on a printer.
50 - 5.346 mg As soon as the result is stable, it is transferred via the interface connection 4 to the printer. Up to the transmission, the triangle symbol is lit up in the display. If the transmission has not yet taken place, you can abort the procedure by pressing the «Print» key again briefly. After the transmission, the result is kept for about 5 seconds in the display for checking purposes, and a ring symbol flashes at the upper right. 5.
Operating instructions METTLER TOLEDO MT/UMT balances The draft shield, the zeroing (taring) of the balance and the print command can also be controlled externally via an electrical contact (relay, proximity switch). The connections for the electrical contacts 3 are at the rear of your balance. 6 Important information regarding your balance 6.
52 - Use the AC adapter holder as a template to mark out the drill holes. - Drill the holes and use the dowels in the installation set supplied should the wall material require it. Fix the AC adapter holder such that the arrow is visible and pointing upward. - Hold the AC adapter so that the connection cable exits from the top and position the four feet in the slots in the holder. Hold the connection cable to the right while doing so and it will slide into the central slot of the holder automatically.
Operating instructions METTLER TOLEDO MT/UMT balances - Reconnect the balance to the power supply. - Depending on the temperature difference between the old and new location, you should allow the balance to acclimatize for 6 to 12 hours before you start weighing again. Set Re-Zero 6.
54 6.3 A well maintained balance is a joy forever - We have enclosed a brush and cleaning tweezers for cleaning your balance. These aids can be found in drawer 10 on the left of the evaluation unit. To clean the weighing chamber, proceed as follows: - Remove the glass cover of the weighing chamber. MT balances: Grasp the weighing chamber plate with the holder at the tip of the cleaning tweezers and lift it together with the weighing pan completely out of the weighing chamber.
Operating instructions METTLER TOLEDO MT/UMT balances Cleaning The balance housing and the weighing pan are made of high-grade, resistant materials. You can thus use any commercially available cleaning agent for cleaning. - 6.
56 6.5 Additional information and accessories for your balance For below-the-balance weighings your balance is equipped with an opening for a hanger. Remove the glass cover of the weighing chamber. Tilt the balance to the side and swing the cover of the opening on the underside of the balance to one side. You can now attach a pan or a holder to the horizontal hole underneath the weighing pan. Do not attach these to either of the two rings! 7 Any questions? 7.
Operating instructions METTLER TOLEDO MT/UMT balances Electrostatic charging of weighing samples and containers Magnetic weighing samples or containers This charging frequently appears in heated rooms with dry air (below 40 45% rel. humidity) and with weighing samples made of glass or plastic. The electrical charges generate forces that can disturb the weighing. The small, circular draft shield with its metal framework and the weighing pan are grounded and considerably reduce the electrostatic charging.
58 Weighing samples or containers that are not at ambient temperature Weighing samples or containers that absorb or evolve moisture readily Weighing samples or containers that are warmer or colder than the ambient temperature of the balance can cause disturbing air currents and buoyancy errors. Weight changes due to the uptake or evolution of surface moisture can also be the cause. These lead to wrong or unstable weighing results.
Operating instructions METTLER TOLEDO MT/UMT balances Contamination 7.2 How can the weighing speed be increased? Powder, liquid or other residues at the edge of the weighing pan or between the weighing pan and the draft cover can lead to unstable displays if the weighing pan no longer has completely unrestricted movement.
60 Readability 7.3 What happens if...? If your application allows, you should reduce the readout increments for the last decimal place. The larger the readout increments (e.g. steps of 5 instead of 1), the more rapid the operation of your balance (see Section 4.4). ...
Operating instructions METTLER TOLEDO MT/UMT balances AUTOCALIN 0.367 - AUTOCALIN ...after connection to the power supply the adjacent display appears? ...during a weighing the audio warning beep sounds and the adjacent symbol appears in the display? -OFF- If the fully automatic self-adjustment is switched on (see Section 4.4), your balance shows that it wishes to adjust itself (AUTOCALIN = Automatic Calibration and Linearization). But there is no need for you to interrupt your work for this.
62 ...the display shows overload or underload? ...the status indicators suddenly disappear? - Overload is displayed when the weight of the loaded weighing sample is larger than the weighing range. - - Underload is displayed when the weighing range is not reached. This applies when the weighing pan is missing, or the weighing pan touches the draft collar and no longer has unrestricted movement.
Operating instructions METTLER TOLEDO MT/UMT balances ...a flashing circle appears above the weight unit? ...no weighing results or nonsensical characters are printed out on the attached printer? 5.346 mg - 5.346 mg To ensure correct functioning of the attached printer, different settings have to be made in the configuration register, including values for the data transmission and the parity (see Section 4.6). Information regarding the correct settings can be found in the handbook of your printer. ..
64 7.4 What does this error message mean? - Error messages in the display draw your attention to faulty operation, namely that the balance was not able to perform a procedure properly. Error messages are supported by an audio signal (“twitter”). no rEF If one of the application units percent (%) or piece (PCS or Stk ) has been selected as weight unit 2, the balance must have a reference weight available. The following errors can appear: No reference weight has been loaded.
Operating instructions METTLER TOLEDO MT/UMT balances The corresponding procedure is aborted automatically and the balance returns to the weighing mode. Fault in the adjustment Err.unStb Err tiME - The adjustment procedure of the balance has been disturbed. - The adjustment is automatically aborted and the balance returns to the weighing mode. 7.5 What does this term mean? We have “translated” the most important technical terms for you.
66 Ambient conditions Local conditions at the location of the balance. The following local conditions can lead to display errors during weighing: unsuitable location, changes in the temperature, atmospheric humidity or the atmospheric pressure, magnetic fields, electrostatic forces, vibrations, drafts caused by air conditioning. Section 7.
Operating instructions METTLER TOLEDO MT/UMT balances Coarse range In comparison with the fine range, in the coarse range you work with a resolution reduced by a factor of 10. This makes the balance faster and the weighing results are extremely stable. Sections 2.3 and 2.4 Configuring Changing the settings in the configuration register. Section 4 Data transmission rate Specifies the rate at which the data are transferred. Also called baud rate: 1 baud = 1 bd = 1 bit/second. Section 4.
68 Digit Display unit; a digit corresponds to a display increment of the last decimal place. - 1 digit = 1d = 0.001 mg in the coarse range - 1 digit = 1d = 0.0001 mg in the fine range See also “Readability” Section 5.5 Dispensing Fine weighing in of powder or small quantities of liquid. Section 3.2 Display Used for communication between balance and user. Shows weighing results, setting options, etc.
Operating instructions METTLER TOLEDO MT/UMT balances Handshake Special control lines or data signals for synchronization of the data flow via the serial interface (feedback between sender and receiver). Section 4.6 Level Device which facilitates the horizontal alignment of the balance. Operates on the principle of a spirit level with an air bubble in a liquid. See also “Leveling”. Section 1.
70 Menu A menu is generally understood to mean a series of selections which the user can match to his needs. In the menu the user can match the balance to the ambient conditions, adjust the balance and do other adjustments. Section 3 and pages 78 and 79 Microbalance Balance with a readability of 1d (1 digit) = 1 µg = 0.001 mg = 0.000001 g. Section 1.1 Overload/underload display Signals the overloading or underloading of the balance through illumination of the horizontal segments in the display. Section 7.
Operating instructions METTLER TOLEDO MT/UMT balances pro FACT Fully automatic adjustment and linearization with two internal weights when necessitated by the operating conditions. proFACT: professional Fully Automatic Calibration Technology. Section 4.4 Reference weight Representative reference weight. Needed among other things for the performance of percent weighings and piece counting.
72 Resolution (resolving power) of a balance See “Readability” Re-Zero International designation for the zeroing of the display and for the zeroing key (the «Re-Zero» key of your balance). Also used for zero setting, taring and the tare bar, although the definitions are not identical. Section 1.3 Sector Part of the configuration register. The configuration register of your balance comprises four sectors, each of which contains various parameters.
Operating instructions METTLER TOLEDO MT/UMT balances Status indicators Symbols that show the most important settings (vibration adapter, weighing process adapter). The symbols can be displayed permanently or for just a few minutes after switching on the balance. Section 4.5 Vibration adapter Possibility to adapt the balance to the local conditions prevailing at the weighing station. Set in the menu. Section 3.
74 Weight unit Unit (e.g. g, mg, ct, etc.) that can be selected in the configuration register. Section 4.5 Zeroing See “Taring” and “Re-Zero”.
Operating instructions METTLER TOLEDO MT/UMT balances 7.6 How are weight units converted? The following table will facilitate the conversion between the weight units available for your balance. Gram g Milligram mg Microgram µg Carat ct/C.M. (metr.) koil Pennyweight dwt Grain GN Momme mo 1g = 1 1’000 1’000’000 5 0.6430149314 15.432358353 0.2666666667 1 mg = 0.001 1 1’000 0.005 0.0006430149 0.0154323584 0.0002666667 1 µg = 0.000001 0.001 1 0.000005 0.0000006430 0.
76 7.7 Where will I find details of...? The following index shows you the page number of these instructions on which you will find the information appropriate to a particular theme. The page numbers in bold print refer to the most important information for the specific search word.
Operating instructions METTLER TOLEDO MT/UMT balances Linearization Self-adjustment, fully automatic 30, 45 Self-test 5, 8, 72 Setting options 13, 21 Stability detection 26, 66, 72 Stabilization time 59, 72 Standby 7, 72 Status indicators 32, 73 Switching off the balance 7, 61 29, 45, 69 Magnetic weighing samples or containers Manual adjustment Menu 57 45, 47 13 Operational setting Overload 23, 25 62 Taring Percent Percent weighing Piece counting Power outage Printer proFACT Program cassette Readabi
78 Press and hold key Weighing mode Menu Configuration register Set reference weight for % 1) Operational setting Basic settings Weight units Interface Set reference weight for piece 1) Return to factory setting Automatic stability detection Weight unit 1 Data transmission mode * * Internal adjustment Data transfer rate * Readability 4) 2) External adjustment 3) * Weight unit 2 Operating instructions METTLER TOLEDO MT/UMT balances *
Readability (fine range) Set weighing process adapter 4) Parity * * * * Set vibration adapter Transmission protocol Autozero AZ Status indicators * * proFACT(Adjustment and linearization) * Fully automatic door function * CALIN Beeper setting End-of-line mode * * * Menu Configuration register 1) Appears only if weight unit 2 is set in the configuration register to percent (%) or piece (Stk or PCS ).
80 Technical Specifications Modell Readability UMT2 UMT5 Comparator MT 5 0.1 µg 2100 mg 0...2100 mg 0.1 µg 5100 mg 0...5100 mg 1 µg 5100 mg 0...5100 mg 0.25 µg 0.25 µg 0.4 µg 0.8 µg 0.9 µg Linearity Linearity referred to 500 mg ±1 µg ±0.5 µg ±4 µg ±2 µg ±4 µg ±2 µg Stabilization time (typical) 10, 16, 24 s 15, 20, 30 s Corresponding to setting of vibration adapter. Built-in adjustment weights 2x1g 2 x 2.5 g 2 x 2.