Operating instructions

27
Operating instructions METTLER TOLEDO MT/UMT balances
Stage Weighing Repeata-
speed bility
off
very fast good
1
2
3*
4
5
6
7 fast very good
*= factory setting
In the
oFF
position, the balance does
not wait for attainment of stability; each
result is considered stable. This setting
is used in special applications in con-
nection with the serial data interface
(see also Section 4.6 “Data transmis-
sion mode”).
d
Readability (selection of the readout
increments)
In this parameter you define the read-
out increments (digits) for the last deci-
mal place. Readout increments of 1, 2,
5 and 10 are available, and these are
symbolized by the following setting
options:
MT5 UMT2 UMT5
0.01 mg* 1 µg* 1 µg*
0.02 mg 2 µg 2 µg
0.05 mg 5 µg 5 µg
0.1 mg 10 µg 10 µg
* = factory setting
Select readout increments as large as
your application allows; the larger the
readout increments, the faster the op-
eration of your balance.