Operating instructions

49
Operating instructions METTLER TOLEDO MT/UMT balances
5.6 How to print out the
weighing result
You can output the weighing result via
the interface connection 4 on a printer.
If you wish to attach a third-party
printer, you match the parameters of
the data interface to the specifications
in the printer handbook (see Sec-
tion 4.6). You will find further details in
the instructions “Bidirectional interface
of the METTLER TOLEDO AT/MT/UMT
balances”, which is also enclosed.
- In the weighing mode press the
«Print» key briefly.
The internal weights are now loaded,
without the balance being adjusted,
and the current, model-dependent ad-
justment value displayed, interrupted
by the display
tESt
.
- When you have read off the value,
abort the test by pressing the «Set»
or «Cancel» key. The balance is
then again in the weighing mode.
Notes:
If the deviation of the adjustment
weight is more than ±0.00015% (1.5
ppm) from the full load, the balance
should be adjusted (see Section 5.4).
If you work with your own, external ad-
justment weights (parameter “adjust-
ment and linearization” set to
CAL
USEr
in the configuration register), you
can check the adjustment of the bal-
ance by loading your own, external ad-
justment weights.
5.000001
5.000000
On/Off
Print
Select 2
METTLER TOLEDO