Operating instructions

50
Operating instructions METTLER TOLEDO MT/UMT balances
- As soon as the result is stable, it is
transferred via the interface connec-
tion 4 to the printer. Up to the trans-
mission, the triangle symbol is lit up
in the display. If the transmission
has not yet taken place, you can
abort the procedure by pressing the
«Print» key again briefly.
After the transmission, the result is
kept for about 5 seconds in the dis-
play for checking purposes, and a
ring symbol flashes at the upper
right.
- Notes:
The type of transmission depends
on the selected data transmission
mode, which you have set in the
fourth sector of the configuration
register (see Section 4.6). If you
have selected the
S.Stb
or
S.Auto
mode and the automatic door func-
tion is switched on, the draft shield
is closed, the result transmitted and
then the draft shield opened again.
If not, you close the draft shield
manually or by pressing the or
key.
- The transmission command can
also be triggered with the hand or
foot switch (available as an acces-
sory).
5.7 You can also operate
your balance by remote
control
The complete operation (keypad and
draft shield) as well as the menu and
the configuration register of your bal-
ance can also be controlled via a com-
puter. This allows you to incorporate
your balance in an automation system
and, for instance, is also suitable for
loading by a robot.