Fault in the adjustment
- The adjustment procedure of the
balance has been disturbed.
- The adjustment is automatically
aborted and the balance returns to
the weighing mode.
The wait time of 1 minute has run
out
This could have the following causes:
- The setting of the automatic stability
detection (ASD) is too high. After
pressing the «Print», «Re-Zero» or
«Set» key (with
%
,
PCS
or
Stk
), sta-
bility is not reached within 1 minute.
- The weight for the external adjust-
ment has not been loaded.
-A wrong weight has been loaded.
The corresponding procedure is
aborted automatically and the balance
returns to the weighing mode.
7.5 What does this term
mean?
We have “translated” the most impor-
tant technical terms for you.
References (in italics) show you the
section(s) in which you can find further
information.
Adjustment
Matching the balance to a reference
weight. Adjustment, see also proFACT
Sections 4.4, 5.4 and 5.5
Adjustment test
Test function which can be called up in
the menu to allow checking of the ad-
justment without adjustment actually
being performed.
Section 5.5
Err.unStb
Err tiME
CAL tESt