Operating instructions
The “Percent Weighing” application
108
9.4.2 Percent Weighing to a Target Value
The “Percent Weighing” application provides you with additional functions to make weighing to a specified target value easier. In the
description which follows, it is assumed that the reference for the percent weighing has already been determined.
Requirements
Before you can input a target value and the associated tolerances, the function keys
shown at left must be activated (section 9.3.2). If you want the specified values to be
shown in the display, you can also activate the information fields with the same name
(section 9.3.3).
Performing a Percent Weighing to a Target Value
Press the function key “Nominal”. Type in the desired target value (e.g. 130 %). Check
the weighing unit which is displayed to the right of the target value. Touching the weigh-
ing unit displays the available units, which include “%” (the “Percent” unit only becomes
available after the reference has been determined).
Note: The units are not converted automatically, so once you have input a value in a
particular unit, the value does not change even if you change the unit.
When you have input the value, press “OK” to activate the target value.
You can use the two function keys “+Tolerance” and “–Tolerance” to specify the accuracy
with which you want to weigh. The input window is similar to the one for the target value.
Both tolerance values are set to 2.5 % at the factory. Instead of a percentage value, you
can also enter a unit of your choice (e.g. “g”). When you have entered the respective value,
press “OK” to activate the tolerance. Percentage weighings which lie outside the tolerances
are specially marked in the report of individual values with “>T” or “<T”.
When you have input the target value and the tolerances, tolerance marks for the graphical
weighing aid (“SmartTrac”) appear in the display. The tolerance marks make weighing
to the target value easier for you. You can weigh your weighing sample roughly until the
lower tolerance value is reached, and then if necessary dispense finely until the target
value is reached.
Performing the Percent Weighing
After you have determined the reference, place the weighing sample on the pan. The
weight of the weighing sample as a percentage of the reference weight appears in the
results display.
Note: If instead of the percentage weight you want to know the absolute weight of the
weighing sample, touch the “%” unit and select the desired weighing unit.
You can print out the result of the percent weighing with the «F» key. You will find an
example of a report in section 9.4.3.