Operating instructions

The “Density” application
114
In the three submenus where you can specify the options for the header line of reports,
the options for reports of individual values, and the options for statistical reports, there
are additional settings available which are described below.
Note: All other items of information on the reports are the same as for the “Weighing”
application (section 5.2.8) and are not listed here.
Report header line
On the second and third pages of this submenu there are further settings for density
determination:
Method”: Reports the selected density determination method.
Aux. Liquid: Reports the selected auxiliary liquid (for density determination
of solid bodies).
Density AL”: Reports the density of the auxiliary liquid which was input using
the function key with the same name. If water or ethanol is be-
ing used, the value taken from the built-in table is reported.
Temp. AL”: Reports the temperature of the auxiliary liquid which was input
using the function key with the same name (for water and
ethanol).
10.3.8 Special report information for density determination
10.3.7 Special information fields for density determination
On the first page of the menu of info fields there are additional settings for density determination:
Method”: Selected type of density determination.
Auxiliary liquid: Selected auxiliary liquid (density determination of solid bod-
ies).
Temp. AL”: Shows the temperature of the auxiliary liquid (distilled water,
ethanol) which was input using the function key with the same
name.
Density AL”: Density of the auxiliary liquid. For water and ethanol, the
density which is automatically read from the built-in density
table is displayed; for other auxiliary liquids the density value
which was input using the function key with the same name
is displayed.
Vol. Sinker”: Volume of the sinker (for density determination of liquids using
a sinker).
Vol. G. Sphere”: Volume of the gamma sphere (for density determination of
pasty substances with the aid of a gamma sphere).
Vol. Pycno”: Volume of the pycnometer (determining the density of liquids
using a pycnometer).
Wgt. Pycno”: Weight of the pycnometer (determining the density of liquids
using a pycnometer).
Factory setting: “Method”, Aux. Liquid”, and “Temp. AL activated (in this or-
der).