Operating instructions
The “Density” application
116
10.3.9 Special SmartSens and ErgoSens settings for density determination
There are additional settings for the SmartSens and ErgoSens sensors for density determination.
“Start” and “Result” emulate the function keys with the same names. “OK” emulates pressing
the key with the same name in the density determination dialogs (but not in the menus)
to confirm inputs and actions.
If one of these settings is activated, the green “F” (function) symbol under the respective
sensor in the status bar lights up.
Factory setting: Model-dependent:
Precision balances: “Off” for all 4 sensors.
Analytical– and
Micro balances:
SmartSens left and right configured for draft shield (doors)
operation.
“Off” for both ErgoSens..
“Vol. Probe”: Reports the volume of the sample (which is calculated by the
firmware).
“Density”: Reports the result of the current density determination.
Factory setting: “Density” and “3 Blank Lines”.
The report of individual values (report of an individual density determination) can be
printed out with the «F» key. You will find an example of a report in section 10.4.6.
Reporting statistical data
On the second and third pages of this submenu you can specify the statistical information
of density determinations that you want to be reported. These settings are only relevant
if you have activated the statistics function (section 10.3.4).
“Method”: Selected density determination method.
“Aux. Liquid”: Selected auxiliary liquid (for density determination of solid
bodies).
“n, x, s, s. rel.”: Number of samples in the current measurement series (“n”),
average (mean) density value of all samples (“x”), and absolute/
relative standard deviation of the current measurement series
(“s”/“s. rel”).
“Min, Max, Diff”: Lowest and highest density values determined in the current
measurement series and the difference between these two
values.
Factory setting: “n, x, s, s.rel” and “Min, Max, Diff”. The settings “Signature”,
“Dash Line” and “3 Blank Lines” are also activated.
The statistics report can be printed out by opening the statistics window and then press-
ing the «F» key. You will find an example of a report and an explanation of the statistics
in section 10.5.