Operating instructions

The “Dynamic Weighing” application
129
With the Dynamic Behavior setting, the balance determines the release of the weighing
result based on your specifications (filter settings):
Stable”: This setting is suitable for relatively stable weighing objects.
Standard”: This setting is suitable for normal weighing objects.
Unstable”: This setting is suitable for unstable weighing objects.
Alternatively, the Time Interval setting allows you to define a fixed amount of time (1
... 99 seconds) until the weighing result is released. This can impair precision, but it can
be practical when the above filter settings result in long measuring times in the case of
extremely unstable weighing materials, or when you have to conclude your measurements
within a certain amount of time.
Factory setting: “Time Interval” = “5 sec”.
The following special function keys are available for dynamic weighing:
Start”: You can use this function key to start a manual weighing cycle. If the Automatic Start” function has been
activated (section 11.3.4), this function key is not required. If the Automatic Start” function has been switched
off, you must activate this function key, otherwise no weighings can be performed!
Result: Opens the results window. This function key is only required if the statistics function has been activated
(section 11.3.11).
CL Result”: Clears the statistics of a weighing series. This key is only required if the statistics function has been acti-
vated.
CL Last”: Deletes the last value which was included in the statistics. This function key is only required if the statistics
function has been activated.
All other function keys are the same as for the “Weighing” application (section 5.2.2).
Factory setting: “ID” and “1/10d” activated (in this order). No specific function keys for dynamic weighing are activated.
11.3.3 Adapting the application to the dynamic behavior of the weighing material
You can use the settings in the “Data Acquisition” menu to adapt the application to the behavior of the weighing object and thereby
optimize the speed of a result determination. The following settings are available:
11.3.2 Special function keys for dynamic weighing