Operating instructions
The “Dynamic Weighing” application
133
11.3.8 Automatic or manual reporting of individual values
In the “Print single value” menu you specify whether the individual values (section 11.3.7) should be printed out automatically or
manually.
“Automatic”: A report of individual values is printed out automatically after
each successfully completed measurement cycle.
“Manual”: As soon as the result of a dynamic weighing is displayed, the
individual value report can be printed out with the «F» key.
Factory setting: “Automatic”
11.3.9 Selecting the destination for the output data
In the “Output channel selection” menu, you can specify the instruments to which the weighing results should be transmitted.
“Printer”: The data are transmitted to the printer according to the selected
protocol settings (section 11.3.7).
“Host”: The weighing results are transmitted to the host computer ac-
cording to the settings for formatting the output data (section
11.3.10).
“Printer + Host”: The data are transmitted both to the printer and to the com-
puter (according to the protocol settings and the settings for
formatting the output data).
Factory setting: “Printer”
Note: If you selected the “Quick procedure” setting for the start of
the weighing (section 11.3.4), only the weight values that are
determined are protocolled on the printer, irrespective of the
protocol settings that were sselected. In the data output to the
host, the settings for formatting the output data are ignored.
The values are protocolled in a fixed format, and the “*” symbol
is inserted before each value.
11.3.10 Formatting the output data
In the “Data output format” menu, you can specify how the weighing results that are
output to the host computer should be formatted. This can be necessary if you operate
your balance together with other instruments, programs, or peripheral devices that require
a particular data format.