11.4.3 Recording statistics of dynamic weighing
Settings
To record statistics of dynamic weighing, the statistics function must be activated (sec-
tion 11.3.11). You should also activate the function keys shown at left (section 11.3.2).
Using the statistics
With the statistics function activated, all weighing results are automatically transferred
to the statistics (up to 999 values). If an incorrect result is included in the statistics by
mistake, you can remove it with the “CL Last” function key. However, you can only remove
the last result that was included. After you have removed the incorrect result, the key
becomes inactive and only functions again after the next result has been transferred to
the statistics.
You can press the “Result” function key to open the statistics window. The key is only
available if there are values in the statistics, otherwise it is shown dimmed and cannot
be operated.
You can print the statistics by pressing the «F» key. You will find explanations of the
individual values in section 11.4.4.
If you want to terminate the current weighing series and clear the statistics ready for a
new series, press the “CL Result” function key. For safety, you will be asked for confirma-
tion before the statistics are finally deleted. Note: If the key is shown dimmed, there are
no values in the statistics.