Operating instructions
The “Differential Weighing” Application
145
Defining the results protocol
In this submenu, you define the format in which the results of differential weighing are
printed. The following settings are available:
“Diff.”: Absolute weight difference between initial weight and residual
weight.
“Diff. %”: Difference between initial weight and residual weight as a
percentage of the initial weight value.
“Abs. %”: Residual weight as a percentage of the initial weight (weigh-in
value).
“Atro AM”: Moisture content of the sample as a percentage of the dry
weight (“ATRO Moisture Content”).
“Atro AD”: Wet weight of the sample as a percentage of the dry weight
(“ATRO Dry Content”).
Factory default setting: “Diff.”, “Diff. %”, “Abs. %” and “Atro AM” (in this order).
To print the results protocol, press the «F» key (prerequisite: the Print key has been
configured for printing series data, see section 12.3.5). Results are printed in the display
unit selected.
The formulae used to calculate the above settings are provided in section 12.6.
Printout footer
On the second page of this submenu, you define which additional information for dif-
ferential weighing is printed in the footer of the printout after the results :
“Series ID”: Prints the series ID.
Factory default setting: “Signature” and “3 Blank Lines” (in this order); no specific
information is activated for differential weighing .
No. of decimal places
In this submenu, you define the number of decimal places to be included in the results.
“1” – “5”: The results are displayed in the printout with the defined number
of decimal places.
Factory setting: “3”
Note: This setting refers only to the results of differential weighing calculated by the
application. Weight values (tare, weigh-in, residual weight) are always recorded in the
maximum resolution of the balance used.
For a sample printout of differential weighing, see section 12.5.5.