Operating instructions

System settings
28
3.9 Settings for the level sensor (Model-dependent)
The built-in level sensor constantly monitors the correct horizontal alignment of your balance. In this menu you can activate or deactivate
the level sensor, and define settings for the output of warnings when incorrect leveling has been detected. You can find information on
the leveling procedure in the Operating Instructions – Part 1. Note: On some models, the level sensor cannot be switched off.
If the level sensor is activated, you can specify whether and how often a warning text
and/or warning beep should occur when the balance is not precisely level by pressing
the Define button.
The following settings are available for the “Warning Text” and “Warning Beep”:
Off”: When the level is incorrect, only the status icon appears in the
upper right-hand corner of the display. There is no warning text
or beep.
Once”: After the level sensor detects that the level is incorrect, the
warning text and warning beep occur only one time.
Repeat”: As long as the balance remains incorrectly leveled, the warning
text and warning beep are repeated every 15 minutes.
Factory setting: Level sensor on/off: Model-dependent.
“Warning Text”: “Once”
“Warning Beep”: “Once”
Note
The Leveling Assistant (see Operating Instructions – Part 1) makes it easier for you to
level the balance. You can call up the Assistant directly from the “Warning Text” window,
(Micro balances from the information field) which is displayed when the leveling is
incorrect. If you deactivate the “Warning text” you have to start the Leveling Assistant
by touching the information eld zone in the display (see Operating Instructions Part
1, section 3.2).
The level sensor is connected to the backlighting of the level indicator. When the level
sensor is switched on, the level indicator is lit up. When the sensor is deactivated, the
illumination of the level indicator is also switched off.