The “Weighing” application
51
Info”: When “Show is pressed, a window opens containing information
about the “MinWeigh” function (method, date of the next test
by the service technician and the minimum required weight
values defined by the service technician in relation to reference
tare values). Press the «F» key to print the information.
Factory setting: “MinWeigh” function deactivated (“Off”).
5.2.15 Defining and activating the tare memory
This menu can be used to define and rename up to 10 tare values, which are available under the “Tare Store” function key (section
5.2.2).
Select the tare memory that you wish to activate/deactivate, rename or modify the weight
value for. Note: Deactivated IDs (“Off”) can no longer be selected under the “Tare Store”
function key.
A window appears, in which you can activate and rename the desired tare memory as
well as modify its weight value.
The designations of the tare memory ex works are “T1” ... “T10”. The names can be cus-
tomized, e.g. the name of a tare container can be used.
To change the current tare memory name, press the corresponding button. An alphanumeric
input field appears, in which the new name can be entered (maximum of 20 characters).
The tare memory is now available with the new name under the “Tare Store” function
key. Note: Even though the maximum permissible length of the name is 20 characters,
we recommend that you use short names, as a maximum of only 10 characters can be
displayed under the “Tare Store” function key.
To enter the weight value, press the corresponding button. A numeric input field appears,
in which the tare value can be entered. Note: Instead of typing in the value, you can
place the relevant tare container on the weighing pan and then press the button with the
balance icon. The measured weight value is thus transferred directly.
Notes on working with the tare memories can be found in section 5.3.2.
Factory setting: All tare memories are deactivated (“Off”)