5.2.17 Settings for SmartSens and ErgoSens
This menu option can be used to activate/deactivate each of the two contact-free sensors (SmartSens) in the top left and right corners
of the terminal and to assign each sensor a function. The same settings are available for a maximum of two external ErgoSens sensors.
The ErgoSens sensor is an optional external sensor with the same functions as the built-in SmartSens sensor. A maximum of 2 external
ErgoSens sensors can be connected to the balance.
Each of the 4 sensors can be assigned one of the following functions by touching the
relevant button:
Off”: The sensor is inactive.
Door”: Model-dependent
Opening and closing of the glass draft shield (doors).
G”: Zero the balance.
H”: Tare the balance.
Print”: Start the print process.
ID1”... ID4”: Open the input window for the relevant ID.
Header”: Print the header.
Footer”: Print the footer.
Pre Tare”: Open the input window to enter a fixed numeric tare value
(tare preset).
1/2d - 1/1000d”: Change the resolution of the weighing result.
AntiStatic Kit”: Model-dependent:
This setting is only available for “ErgoSens 1” and “Ergo
Sens 2”. The optional antistatic kit is connected to one of the
two ErgoSens connections (“Aux 1”/“Aux 2”). For the antistatic
kit to function, it must be selected under “ErgoSens 1” or “Er-
goSens 2” depending on the connection used. You can find
additional information on the antistatic kit in section 5.2.18..
Transfer”: Transmits the stable formatted weight value across the inter-
face.
5.2.16 Settings for the automatic tare function
This menu option can be used to specify whether and under which conditions the balance is to automatically interpret the first weight
following a zero setting as the tare. Notes on working with the automatic tare function can be found in section 5.3.2.
When you activate the AutoTare function (“On”), press the button to define the weight
criterion for the automatic tare function:
Limit”: This value specifies the minimum weight that must be placed
on the weighing pan for it to be automatically saved as the tare.
If the weight is below the limit, the value is not automatically
transferred to the tare memory.
Note: Instead of typing in the value, you can place your lightest
tare container on the weighing pan and then press the but-
ton with the balance icon. The measured weight value is thus
transferred directly as the limit.
Factory setting: “Off” (automatic tare function is deactivated).