Operating instructions

The “Weighing” application
53
5.2.18 Settings for the optional antistatic Kit (Model-dependent)
Settings for the optional antistatic kit eliminates the static charge of weighing material. The antistatic kit is connected to one of the
two Aux 1” or Aux 2” connections at the rear of the balance (both of these connections are also used for the optional ErgoSens). The
antistatic kit must be selected in the ErgoSens settings for the balance to recognize it (see section 5.2.17; also see the instruc-
tions supplied with the antistatic kit).
In this menu option you can make settings for the operating principle of the antistatic
kit.
The following settings are available:
After door opened”: Here you can define whether ionization will occur when the
glass draft shield (doors) is opened (“Active”) or whether no
ionization will occur (“Inactive”). With the Active” setting,
ionization will occur continuously until the draft shield (doors)
is closed again (a maximum of 10 minutes).
After door closed”: This setting defines the number of seconds ionization should
occur after the draft shield (doors) close (adjustment range
0 ... 1000 seconds). Note: If you set it for “0” seconds, there
will be no ionization after the draft shield (doors) close
Factory setting: Ionization occurs when the draft shield is open (“Active”).
Ionization occurs for 5 seconds after the draft shield (doors)
close.
Note: If a function is activated, which emulates a key on the terminal, the relevant icon lights up below the corresponding sensor in the
status bar (K, “ G”, H oder F”). For all other settings (that emulate function keys of the same name), the green F (function)
icon lights up. No icon lights up when the sensor is deactivated.
Factory setting Model-dependent: Precision balances: “Off” for all 4 sensors.
Analytical- and
Micro balances:
SmartSens left and right configured for draft shield (doors) opera-
tion. “Off” for both ErgoSens.