Operating instructions
The “Weighing” application
59
Pressing this function key triggers a balance adjustment using the built-in calibration
weight. This can be implemented at any time.
You can hear how the internal weight is placed on the pan and then lifted off again in
a motorized process. During adjustment the same window is displayed as for ProFACT
adjustment (see section 3.4 of the separate document Operating Manual for XP Bal-
ances – Part 3).
Following the adjustment operation, one of the following messages is displayed:
Adjustment done. Press “OK” to return to the application.
If a printer is connected to the balance, the adjustment is printed automatically according
to the specifications selected in the system settings for printing adjustments (section 3.8
of the separate document Operating Manual for XP Balances – Part 3). A sample printout
can be found in section 5.4.5.
Adjustment abort. This message also appears when you cancel adjustment. You can
repeat the adjustment operation or press “OK” to return to the application.
Note: The procedure described corresponds to the factory setting. Advanced Options can
be used to extend the internal adjustment procedure with internal tests (see section 3.4.1
of the separate document Operating Instructions for XP – Part 3.
5.4 Adjusting the balance and testing adjustments
Your balance is set to ProFACT (see Operating Instructions – Part 3) fully automatic adjustment ex works. ProFACT automatically
adjusts the balance according to preselected criteria. However, the balance can be adjusted manually and/or tested using the internal
weight or an external weight at any time.
In the following descriptions, it is assumed that the relevant function keys for adjustment and tests (section 5.2.2) are active.
5.4.1 Adjustment using the internal weight
5.4.2 Adjustment using an external weight
Pressing this function key triggers a balance adjustment using an external calibration
weight. This can be implemented at any time, but you can also specify in the system set-
tings that the balance issues a calibration reminder on specific days of the week and at
a specific time (section 3.5 of the separate document Operating Manual for XP Balances
– Part 3). Note: Depending on country-specific regulations, adjustment using an external
weight may not be available on certified balances.