Operating instructions

The “Weighing” application
60
For automatic external adjustment only: If you specify in the system settings that the
balance should automatically request adjustment (section 3.5 of the separate document
Operating Manual for XP Balances – Part 3), the window opposite will appear at the se-
lected time. If you want to carry out the adjustment, press Yes to start the adjustment
process described below. If, on the other hand, you press Later”, you will be prompted to
carry out the adjustment again after 15 minutes. The adjustment weight selected last is
always used for automatic external adjustment. Note: The small weight icon for automatic
external adjustment in the upper right of the display is deleted after adjustment is suc-
cessfully completed or if, at the time of the second request, you indicate your decision
not to carry out the adjustment (“No”).
A list for selecting the adjustment weight will appear when you press the function key
Adjust. ext.”. Note: All adjustment weights in the figure shown here still have the names
they were given at the factory, with the exception of the first weight, which was assigned
a unique name (section 3.3.1 of the separate document Operating Manual for XP Bal-
ances – Part 3). The adjustment starts immediately after the weight is selected
After the adjustment process starts, you will be prompted to place the adjustment weight.
If available, the ID and the certificate number of the corresponding weight are displayed.
The required weight value flashes at the bottom of the window.
Important: Make sure you place the correct weight on the pan, otherwise the adjustment
operation will be stopped and aborted with an error message after a specific period of
time has elapsed. The adjustment weight can be defined in the system settings (section
3.3.1 of the separate document Operating Manual for XP Balances – Part 3).
Note: Press Cancel to exit the current adjustment operation at any time.
Once the requested weight is placed on the pan, the adjustment operation is carried
out automatically. At the end of the adjustment operation you are prompted to lift off the
weight. Remove the adjustment weight from the weighing pan.
The balance confirms that adjustment was carried out successfully. Press OK to return
to the application.
If a printer is connected to the balance, the adjustment is printed automatically according
to the specifications selected in the system settings for printing adjustments (section 3.8
of the separate document Operating Manual for XP Balances Part 3). A sample printout
can be found in section 5.4.5.
In the event of an error during adjustment, an error message is displayed, which is the
same as the message for adjustment using the internal weight (see previous section).