Operating instructions
The “Statistics” application
66
“Diff”: Difference between the smallest and largest weight value.
“Nominal”: Shows the nominal weight entered via the function key of the
same name.
“+Tol” and “-Tol”: These information fields show the tolerances entered via the
function keys of the same name.
All other information fields are identical to those for the “Weighing” application (section
5.2.4).
Factory setting: “n”, “x” and “s” are activated (in this order).
The three submenus for defining the record header, printing single values and for the result
contain additional settings for the statistics, which are described below.
Note: The other available record information is the same as for the “Weighing” application
(section 5.2.8) and is not described here.
Record header
Additional settings for the statistics are available on the second page of this submenu:
“Plausibility”: The defined limit for plausible measured values is printed
(section 6.3.7).
“Max n”: The defined maximum number of samples in the series is
printed.
“Nom., +/-Tol”: Prints the defined nominal weight and the defined toler-
ances.
Factory setting: “Appl. Name” (“Statistics” is printed) and “Date/Time” (in this
order); no specific information is activated for the statistics.
6.3.5 Special record information for the statistics
6.3.4 Specifications for automatic weight transfer
This menu option can be used to specify whether and under which conditions the balance is to automatically transfer a stable weight value
to the statistics (this means you do not have to press the “M+” function key). In addition, the weight value is printed automatically.
When you activate this function (“On”), press “Define” to define the criteria for automatic
weight transfer:
“Limit”: This value specifies the weight limit that must be exceeded to
trigger automatic weight transfer.
“Delay Time”: As soon as the weight limit is exceeded, the “Delay Time” starts.
Once this time has elapsed, the weight value is recorded and
transferred to the statistics or transmitted via the interface.
Factory setting: “Off” (automatic weight transfer is deactivated).