Operating instructions

The “Formulation” application
78
Net %”: Prints the weight of the current component as a % of the
nominal weight.
Diff.”: Prints the difference between the actual and nominal weight
of the current component.
Diff. %”: Prints the percentage deviation of the current component
between its actual weight and nominal weight.
Factory setting: “Net”; no specific information is activated for the Formula-
tion.
Note: The names of the four IDs are specified ex works but can be changed (section
7.3.6).
A single value is printed automatically when the “M+ function key is pressed during
formulation. A single value can also be printed separately by pressing the «F» key.
Printing the result
The second and third page of this submenu can be used to define which additional
formulation information is to be included in the printed results record:
Recipe Name”: Prints the name of the current formula.
Recipe ID”: Prints the identification (ID1) of the current formula.
Nominal Tot.”: Prints the sum of the nominal weight values of all components
of the current formula.
Num. of Comp.”: Prints the number of components of the current formula.
Comp. Counter”: Prints the current state of the component counter (number of
the current component).
>Tol+, <Tol-”: Prints the number of components that are outside the upper
or lower weight tolerance.
Net Tot.”: Prints the net total weight of all components weighed in.
Factory setting: “Tare”, “Net Tot.”, “Gross”, “Signature”, “Dash Line” and “3 Blank
Lines” (in this order).
The results record is printed by pressing the «F» key when the results window is open, or
printed automatically once the last component of a formula has been transferred.
A sample printout for formulation can be found in section 7.6.5.