Operating instructions

The “Piece Counting” application
94
Min and Max”: Smallest and largest recorded piece counts in a series of piece
counts.
Diff”: Difference between the largest and smallest piece counts in
a series of piece counts.
Nominal”: Shows the nominal piece count entered via the function key
of the same name.
+Tol and -Tol”: These information fields show the tolerances entered via the
function keys of the same name.
All other information fields are identical to those for the “Weighing” application (section
5.2.4).
Factory setting: “PcsWgt” is activated.
8.3.5 Specifications for automatic weight transfer
This menu option can be used to specify whether and under which conditions the balance is to automatically transfer a stable weight
value to the statistics (this prevents you from having to press the M+ function key during series counting). Also, the piece count is not
automatically printed.
When you activate this function (“On”), press “Define” to define the criteria for automatic
weight transfer:
Limit”: This value defines the smallest modification (piece count
or weight, depending on the unit selected) required to trig-
ger an automatic transfer of the measurement value to the
memory.
Delay Time: As soon as the minimum weight change is reached, the “Delay
Time” starts. Once this time has elapsed, the weight value is
recorded and transferred to the statistics or transmitted via the
interface.
Factory setting: “Off” (automatic transfer is deactivated).
8.3.6 Additional unit for piece counting
In addition to the common weighing units, the menus for the “Display Unit” and the “Info
Unit” also contain the “PCS” unit (provided that a reference piece weight was already
determined).
Note: The “PCS” unit does not have to be explicitly selected for piece counting, since the
display unit always switches to “PCS” automatically when the reference piece weight is
being determined. The desired unit can be selected at any time after the reference piece
weight is determined (unless a value has already been transferred to the memory during
a series of piece counts; in this case, switching between “PCS” and the other weighing
units is not possible until the results have been cleared).
Factory setting: “g” (gram) for “Display Unit” and “Info Unit”.