Operating instructions

The “Piece Counting” application
96
Printing the result
The second and third page of this submenu can be used to define which additional
information for piece counting is to be included in the printed results record:
Max n”: Defined maximum number of piece counts in the series.
Nom.,+Tol,-Tol”: Defined nominal piece count and selected tolerances.
>Tol+, <Tol-”: Number of piece counts outside the upper and lower tolerance
range.
n”: Number of completed piece counts in a series.
x”: Average piece count of all counts in a series.
s and s.rel”: Standard deviation as an absolute and percentual value.
Note: These two values are printed only if the memory contains
at least 3 values. Otherwise, horizontal dashes appear in place
of the values.
Min, Max, Diff.”: Smallest and largest recorded piece count within the current
series of measurements, and the difference between the largest
and smallest piece count.
Sum”: Total of the values of all saved single measurements in the
current display unit.
SumPcs”: Total piece count of all saved single counts.
Factory setting: “n”, “x”, “s”, “s.rel”, “Min, Max, Diff.”, “Sum” and “SumPcs”, as
well as “Signature” and “3 Blank Lines” are activated (in this
order).
The results record is printed by pressing the «F» key when the results window is open. If
a specific number of single counts is defined for a series of piece counts (“Max n”), the
results record is printed automatically as soon as the result of the last count has been
transferred to the statistics.
A sample printout for a piece count can be found in section 8.4.5.
8.3.8 Special SmartSens and ErgoSens settings for piece counting
Additional settings for piece counting are available for the SmartSens and ErgoSens sensors.
“FixPcs”, “Result” and “M+” emulate the function keys of the same name. “OK” emulates
pressing the button of the same name in the piece counting dialog boxes (but not in the
menus) to confirm entries and actions.
If one of the above settings is activated, the green F (Function) icon lights up below the
corresponding sensor in the status bar.
Factory setting: Model-dependent
Precision balances: “Off” for all 4 sensors.
Analytical– and
Micro balances:
SmartSens left and right configured for draft shield (doors)
operation.
“Off” for both ErgoSens..