Operating instructions

68
Operating instructions METTLER TOLEDO AT balances
Automatic door function
Automatic opening and closing mecha-
nism for the glass draft shield that
ensures that the balance and the draft
shield are always in the correct position
for the next weighing step.
Section 3.4
Automatic stability detection
Continuous automatic check of the
weight value for stability. The weight
display is considered stable when its
change with time lies within a preset
limit. Different stages can be configured
by the user and correspond to various
limits. The stability of the display is
attained when the stability detection
symbol in the upper left corner of the
display fades.
Also with Re-Zero (taring) the display is
set to zero only when stability has been
achieved.
Section 4.4
Ambient conditions
Local conditions at the location of the
balance. The following local conditions
can lead to display errors during
weighing: unsuitable location, changes
in the temperature, atmospheric humi-
dity or the atmospheric pressure,
magnetic fields, electrostatic forces,
vibrations, drafts caused by air
conditioning.
Section 7.1
Analytical balance
Balance with a readability of 1 d
(1 digit) = 0.1 mg.
Analytical range
Weighing range with a readability of 1 d
(1 digit) = 0.1 mg.
Section 2.4
Application unit
Unit selectable in the configuration
register that is used for applications in
which not only the weight, but also a
quantity derived from it (piece, percent)
is determined and displayed.
Sections 4.5, 5.2 and 5.3