User manual

OperationIND445
15User manual 22013194B 05/08
2.7 Weighing-in to a target weight and checkweighing
The terminal IND445 allows the weighing-in of goods to a particular target weight
within defined tolerances. With this function it is possible to check whether weighed
materials are within a defined tolerance range.
The terminal IND445 has a total of 100 memory locations for frequently used tare val-
ues, average piece weights, target weights and target quantities. In the factory setting,
the memory locations 81 to 90 are reserved for target weights. The saved target
weights are also preserved when the terminal is switched off.
2.7.1 Saving target weights
1. Enter the memory location number (factory setting: 81 ... 90) and keep
pressed until the confirmation tArGEt appears in the display.
2. Enter the target weight in the defined unit, e.g. 1.5 kg, and confirm with .
The display tOLER appears and + flashes.
3. Enter the upper tolerance in the displayed weight unit, e.g. 0.1 kg, and confirm with
:
-or-
Press , enter the upper tolerance range in per cent and confirm with .
The display tOLEr appears and flashes.
4. Enter the lower tolerance accordingly.
The scale returns to weighing mode.
Note If a target weight had already been saved under the selected memory location, the
message rEPLACE appears in the display.
To save the new target weight, press . The old target weight is overwritten.
To abort the save process, press . The previous memory location assignment
remains valid.
2.7.2 Calling up target weights
Enter the number of the memory location with the required target weight (factory
setting: 81 ... 90) and press briefly.
The selected target weight and the tolerances are loaded from the memory and
appear briefly in the display. The scale is now ready for weighing-in or check-
weighing.
T