Interface description
User manual 22013194B 05/08
5.2 TOLEDO Continuous mode
5.2.1 TOLEDO Continuous commands
In TOLEDO Continuous mode the scale supports the following input commands:
5.2.2 Output format in TOLEDO Continuous mode
Weight values are always transferred in TOLEDO Continuous mode in the following
format:
Command Meaning
P Printing out the current result
T Taring of the scale
Z Zero setting of the display
C Deleting of the current value
S Determining the reference
Status Field 1 Field 2
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
STXSWASWBSWCMSD – – – –LSDMSD – – – –LSDCRCHK
Field 1
Field 2
STX
SWA, SWB, SWC
MSD
LSD
CR
CHK
Cont-Wt: 6 digits for the weight value that is transferred without comma and unit
Cont-Wt: 6 digits for the tare weight that is transferred without comma and unit
ASCII character 02 hex, character for "start of text"
Status words A, B, C, see below
Most significant digit
Least significant digit
Carriage Return, ASCII character 0D hex
Checksum (2-complement of the binary sum of the 7 lower bits of all the characters sent
beforehand incl. STX and CR)