Event and error messages
User manual 22013194B 05/08
6 Event and error messages
Error Cause Remedy
Display Dark
Back lighting set too dark Set back lighting (b.LIGHt)
brighter
No mains voltage Check mains
Unit switched off Switch on unit
Mains cable not plugged in Plug in mains plug
Brief fault Switch device off and back on
again
Insufficient load
Load plate not on the scale Place load plate on the scale
Set to zero
Weighing range not reached
Overload
Weighing range exceeded Unload scale
Reduce preload
Result not yet stable If necessary adjust vibration
adapter or weigh dynamically
Function not permissible Unload scale and set to zero
Zeroing not possible with over-
load or insufficient load
Unload scale
Reference weight too low Select and place larger number of
reference parts on the scale
No valid value from the reference
scale
Check cable connection between
the units
Check interface settings
No calibration Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
Calibrate scale
Call METTLER TOLEDO Service
Average piece weight too low Counting is not possible on this
scale with this average piece
weight