User manual

Event and error messagesIND445
53User manual 22013194B 05/08
Unstable weight value when
referencing
Ensure stable surroundings
Ensure that the weighing pan is
freely movable
Adjust vibration adapter
Impermissible target value or
impermissible tolerance
Repeat input with permissible
values
Setting the average piece weight
impermissible during weight
accumulating
End weight accumulating
Reset average piece weight
Switching the weighing unit
impermissible during weight
accumulating
End weight accumulating
Switch weighing unit
Printout not yet ended End printout
Repeat required action
Switching the weighing unit
impermissible during dynamic
weighing
End dynamic weighing
Switch weighing unit
EAROM checksum error Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
Call METTLER TOLEDO Service
Reference optimization not possible
The total weight of the reference
parts exceeds 4 % of the scale
capacity
Put on fewer reference parts
No additional parts were put on
the scale for manual reference
optimization
Put on reference parts for optimi-
zation
Call METTLER TOLEDO Service
Weight display unstable
Restless installation location Adjust vibration adapter
Draft Avoid drafts
Restless weighing sample Dynamic weighing
Contact between weighing pan
and/or weighing sample and
surroundings
Remedy contact
Mains fault Check mains
Error Cause Remedy