User manual METTLER TOLEDO Compact scales BBA422 / BBA425 / BBK422 www.mt.
@@@@@
BBA422 / BBA425 / BBK422 Table of contents Table of contents Page User manual 22011381A 03/05 1 1.1 1.2 1.3 Introduction .....................................................................................4 Safety instructions .............................................................................4 Description ......................................................................................5 Putting into operation ........................................................................
BBA422 / BBA425 / BBK422 Introduction 1 Introduction 1.1 Safety instructions CAUTION! Do not use BBA422 / BBA425 / BBK422 in hazardous areas! Our product range includes special devices for hazardous areas. CAUTION! Use only scales with Protection Class IP65, if: • the scale is used in wet areas • wet cleaning is necessary • the scale is used in a dusty environment Even with Protection Class IP65, the scale must not be used in environments with corrosion risk.
BBA422 / BBA425 / BBK422 Introduction CAUTION! Handle the compact scale with care. The scale is a precision instrument. ▲ When the weighing pan has been removed, never clean the area under the load plate holder with a solid object! ▲ Do not put excessive loads on the scale. ▲ Avoid banging the weighing pan. Disposal ➜ Observe the valid environmental regulations when disposing of the scale.
BBA422 / BBA425 / BBK422 Introduction 1.2.
BBA422 / BBA425 / BBK422 1.2.2 Introduction Display 1 Active interface 2 Weighing range display 3 Battery charge level; only present on scales with a battery 4 Symbol for displaying net values 5 Symbol for dynamic weighing 6 Weight units 7 7-segment display, 7 digits, with decimal point 8 Stability monitor (goes out when a stable weight value is reached) 9 Sign 10 Identification for changed or calculated weight values, e. g.
BBA422 / BBA425 / BBK422 Introduction 1.2.
BBA422 / BBA425 / BBK422 1.3 1.3.1 Introduction Putting into operation Selecting or changing the location The correct location is crucial to the accuracy of the weighing results! ➜ Select a stable, vibration-free and if possible a horizontal location. The ground must be able to safely bear the weight of the fully loaded scale.
BBA422 / BBA425 / BBK422 Introduction 1.3.2 Connecting the power supply CAUTION! Before connecting the scale to the mains, check whether the voltage value printed on the rating plate corresponds with the local mains voltage. ▲ Never connect the device if the voltage value printed on the rating plate is different to the local mains voltage. ➜ Plug the mains plug into the socket. After connection, the device performs a self-test. When the zero display appears, the device is ready to weigh.
BBA422 / BBA425 / BBK422 2 Operation Operation 2.1 Switching on Switching on and off ➜ Press . The scale conducts a display test. When the weight display appears, the scale is ready to weigh. Switching off ➜ Press . Before the display goes out, -OFF- appears briefly. 2.2 Zeroing / Zero point correction Zeroing corrects the influence of slight changes on the load plate. Manual 1. Unload scale. 2. Press . The zero display appears.
BBA422 / BBA425 / BBK422 Operation 2.4 2.4.1 Weighing with tare Taring ➜ Place the empty container on the scale and press T . The zero display and the symbol NET appear. The tare weight remains saved until it is cleared. 2.4.2 Clearing the tare ➜ Unload scale and press T . The symbol NET goes out, the zero display appears. or ➜ Press C . The symbol NET goes out, the gross weight appears in the display. If A.
BBA422 / BBA425 / BBK422 2.5 Operation Dynamic weighing With the dynamic weighing function, it is possible to weigh restless weighing samples such as live animals. If this function is activated, the symbol appears in the display. With dynamic weighing, the scale calculates the mean value from 56 weighing operations within 4 seconds. With manual start Prerequisite AVErAGE -> MAnuAL is selected in the menu. The weighing sample must be heavier than 5 scale divisions. 1.
BBA422 / BBA425 / BBK422 Operation 2.7 Cleaning CAUTION! Electric shock hazard! ▲ Before cleaning with a damp cloth, pull out the mains plug to disconnect the unit from the power supply. CAUTION! When the weighing pan has been removed, never clean the area under the load plate holder with a solid object! This could damage the weighing cell. Other cleaning information: • Use damp cloths. • Do not use any acids, alkalis or strong solvents.
BBA422 / BBA425 / BBK422 3 Settings in the menu Settings in the menu Settings can be changed and functions can be activated in the menu. This enables adaptation to individual weighing requirements. The menu consists of 6 main blocks containing various submenus on several levels. 3.1 3.1.1 Operating the menu Calling up the menu and entering the password The menu differentiates between 2 operating levels: Operator and Supervisor. The Supervisor level can be protected by a password.
BBA422 / BBA425 / BBK422 Settings in the menu 3.1.2 Selecting and setting parameters COdE tErMINL APPLIC SCALE dISPLAY T T tArE ... dEVICE ACCESS SLEEP COM 1 rESEt PWr.OFF b.LIGHt T ➜ Scroll back: Press . ➜ Press . Exiting menu 1. Press . COM 2 ... T ➜ Scroll forward: Press Activating menu items/ accepting selection End T T T Scrolling on one level dIAGNOS T T CAL COMMUNI T T MOdE PriNtEr PArAMEt rSt.COMX . The last menu item END appears. 2. Press .
BBA422 / BBA425 / BBK422 3.2 Level 1 Level 2 SCALE CAL Settings in the menu Overview dISPLAY Level 3 Level 4 Level 5 Level 6 Page 19 UNIt1 g, kg, oz, lb, t UNIt2 g, kg, oz, lb, t 19 rESOLU UNt.rOLL ON, OFF A-tArE ON, OFF ChAIn.tr ON, OFF A.CL-tr ON, OFF ZErO AZM OFF; 0.5 d; 1 d; 2 d; 5 d; 10 d rEStArt ON/OFF FILtEr VibrAt LOW, MEd, HIGH, PrOCESS UNIVEr, dOSING StABILI FASt, StAndrd, PrECISE FACt tEMP OFF, 1K, 2K, 3K, 5K 20 Min.
BBA422 / BBA425 / BBK422 Settings in the menu Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 dt-G GrOSS ON, OFF tArE ON, OFF nEt ON, OFF Page COnt-Wt 2nd.dISP PriNtEr PArAMEt COMMUNI dIAGNOS OPtION tEmPLat StdArd, tEMPLt1, tEMPLt2 ASCi.Fmt LINE.FMt MULtI SINGLE LENGtH 1 ... 100 SEPArAt , ;... Add LF 0 ... 9 bAUd 300 ... 38400 PAritY 7 nonE, 8 nonE, 7 odd, 8 odd, 7 EVEN, 8 EVEN H.SHAKE NO, XONXOFF, nEt 422, nEt 485 NEt.Addr 0 ...
BBA422 / BBA425 / BBK422 Settings in the menu 3.3 Scale settings (SCALE) 3.3.1 CAL – calibration (adjustment) This menu item is not available for certified scales without internal calibration weight. Internal For scales with an internal calibration weight: 1. Unload scale. 2. Activate menu item CAL with . The scale calibrates with the internal calibration weight. -Int CAL- appears in the display.
BBA422 / BBA425 / BBK422 Settings in the menu 3.3.4 ZERO – automatic zero update On certified scales, this menu item does not appear. AZM Switching on/off automatic zero update and selecting zeroing range. Possible settings: OFF; 0.5 d; 1 d; 2 d; 5 d; 10 d 3.3.5 RESTART – automatic saving of zero point and tare value When the Restart function is activated, the last zero point and tare value are saved.
BBA422 / BBA425 / BBK422 3.3.8 Settings in the menu MIN.WEIG – minimum weight This menu item appears only if the service technician has saved a minimum weight. Switching minimum weight function on/off ON/OFF If the weight on the scale falls below the stored minimum weight, an * appears on the display in front of the weight indicator. 3.3.
BBA422 / BBA425 / BBK422 Settings in the menu 3.5 3.5.1 Terminal settings (TERMINAL) DEVICE – Sleep mode, energy-saving mode and display backlighting This menu item only appears on devices in mains operation. SLEEP When SLEEP is activated, the scale switches off display and backlighting after the time period set when not in use. The display and backlighting are switched on again at the press of a key or if the weight changes.
BBA422 / BBA425 / BBK422 3.6 3.6.1 Settings in the menu Configuring interfaces (COMMUNICATION) COM1/COM2 -> MODE – operating mode of the serial interface Print Manual data output to the printer with A.Print Automatic output of stable results to the printer (e. g. for series weighing operations) CONtINU Ongoing output of all weight values via the interface dIALOG Bi-directional communication via MT-SICS commands, control of the scale via PC CONt.
BBA422 / BBA425 / BBK422 Settings in the menu NET.Addr Assigning network address: 0 ... 31, only for NET 485 ChECSuM Activating checksum byte (appears only in TOLEDO Continuous mode) Vcc Switching 5V voltage, e.g. for a bar code reader, on / off 3.6.4 COM1/COM2 -> RESET COM1/RESET COM2 – resetting serial interface to factory settings Confirmation inquiry SUrE? • Reset interface settings to factory settings with • Do not reset the interface settings with 3.6.
BBA422 / BBA425 / BBK422 3.6.6 tEMPLt1/tEMPLt2 LINE 1 ... 20 Settings in the menu DEF.PRN – configuring templates Selecting Template 1 or Template 2 Select line NOt.USEd • Line not used HEAdEr • Line as header. The contents of the header must be defined via an interface com- mand, see Section 4.1.
BBA422 / BBA425 / BBK422 Settings in the menu 3.7 Diagnosis and printing out of the menu settings (DIAGNOS) tESt SC Internal Testing scale with internal calibration weight • -Int CAL- appears in the display during the test. • After completion of the test, ideally *d=0.0g briefly appears in the display, after which the scale changes to the next menu item KboArd. External Testing scale with external calibration weight 1. The scale checks the zero point. -0- appears in the display.
BBA422 / BBA425 / BBK422 4 Interface description Interface description 4.1 SICS interface commands The compact scales BBA422 / BBA425 / BBK422 support the command set MT-SICS (METTLER TOLEDO Standard Interface Command Set). With SICS commands, it is possible to configure, query and operate the scales from a PC. SICS commands are divided up into various levels. For further information about the MT-SICS commands set, see MT-SICS Manual (Order No. 22 011 459) or contact METTLER TOLEDO Customer Service. 4.
BBA422 / BBA425 / BBK422 Interface description Command Meaning P101 Print out stable weight value P102 Print out current weight value immediately PWR Power On/Off SIRU Send weight value in the current unit immediately and repeat SIU Send weight value in the current unit immediately SNR Send stable weight value and repeat after every weight change SNRU Send stable weight value in the current unit and repeat after every weight change SRU Send weight value in the current unit and repeat ST
BBA422 / BBA425 / BBK422 4.1.2 Interface description Requirements for communication between scale and PC • The scale must be connected to the RS232, RS485, USB or Ethernet interface of a PC with a suitable cable. • The interface of the scale must be set to "Dialog" mode, see Section 3.6.1. • A terminal progam must be available on the PC, e.g. HyperTerminal. • The communication parameters baud rate and parity must be set in the terminal program and on the scale to the same values, see Section 3.6.3. 4.
BBA422 / BBA425 / BBK422 Interface description 4.2 4.2.1 TOLEDO Continuous mode TOLEDO Continuous commands The scale supports the following input commands in TOLEDO Continuous mode: 4.2.2 Command Meaning P Print out the current result T Tare the scale Z Zero the display C Clear the current value Tx.
BBA422 / BBA425 / BBK422 Interface description Status word A Status Bit Function Selection 6 5 Decimal position X00 0 1 4 3 2 1 0 0 0 0 X0 0 0 1 X 0 1 0 0.X 0 1 1 0.0X 1 0 0 0.00X 1 0 1 0.000X 1 1 0 0.
BBA422 / BBA425 / BBK422 Event and error messages 5 Event and error messages Error Cause Remedy Display Dark • Back lighting set too dark ➜ Set back lighting (b.
BBA422 / BBA425 / BBK422 Event and error messages Error Cause Remedy Weight display unstable • Restless installation location ➜ Adjust vibration adapter • Draft ➜ Avoid drafts • Restless weighing sample ➜ Dynamic weighing • Contact between weighing pan ➜ Remedy contact and/or weighing sample and surroundings Incorrect weight display • Mains fault ➜ Check mains • Incorrect zeroing ➜ Unload scale, set to zero and repeat weighing operation • Incorrect tare value ➜ Clear tare • Contact betw
BBA422 / BBA425 / BBK422 Technical data and accessories 6 Technical data and accessories 6.1 6.1.1 Technical data Type key The compact scales BBA422 / BBA425 / BBK422 are available with various capacities and platforms that can be seen from the complete type designation. Example 6.1.
BBA422 / BBA425 / BBK422 Technical data and accessories BBA422 / BBA425 / BBK422 Mains connection Direct connection to the mains (MAINS supply voltage fluctuations up to ±10% of the nominal voltage): • 230 V, 50 Hz, 70 mA • 240 V, 50 Hz, 70 mA • 120 V, 60 Hz, 90 mA • 100 V, 50/60 Hz, 90 mA For battery operation: • Connection via mains adapter: 90 – 264 V, 47 – 63 Hz, 300 mA • Infeed on the unit: 24 V, 1.
BBA422 / BBA425 / BBK422 Technical data and accessories 6.1.3 Weighing ranges and readability BBA4.. The compact scales BBA4.. with strain gauge weighing cells are supplied in the configuration 2 x 3000 d. Higher legibilities are available from the factory with the optional "Premium" weighing cells. Capacity 6.1.4 Configuration 2 x 3000 d (standard) 1 x 6000 d (with optional "Premium" weighing cells Weighing ranges Readability (certified) Weighing range Readability (certified) 3 kg 1.
BBA422 / BBA425 / BBK422 Technical data and accessories 6.1.5 Dimensions B D E C A K F H I L G A B C D E F G H I K L 1) 335 265 100 240 200 46 276 208 216 165 165 1) 335 265 100 240 200 46 276 208 216 – – 370 360 115 350 240 52 310 304 310 – – XS SM 1) LA 1) Dimenions in mm 6.1.6 User manual 22011381A 03/05 Net weights Model without battery with battery with internal calibration weight (without battery) BBA422 – .. SM 4.6 kg 5.
BBA422 / BBA425 / BBK422 Technical data and accessories 6.1.7 Interface connections The compact scale can be fitted with a maximum of 2 interfaces. The following combinations are possible: 6.1.8 COM1 COM2 RS232 – RS232 RS232 RS485 RS232 COM1 can be optionally operated as RS422 or RS485 RS232 Ethernet not for BBA425 RS232 USB RS232 Digital I/O Assignment of the interface connections Pin 6.
BBA422 / BBA425 / BBK422 7 Appendix Appendix 7.1 Information for certified scales in EC countries M Weighing instruments verified at the place of manufacture bear the preceding mark on the packing label and green "M" sticker on the descriptive plate. They may be set to work immediately. M Weighing instruments which are verified in two steps have no green "M" on the descriptive plate and bear the preceding identifaction mark on the packing label.
BBA422 / BBA425 / BBK422 Appendix 7.3 Table of Geo Values For weighing instruments verified at the manufacturer’s, the geo value indicates the country or geographical zone for which the instrument is verified. The geo value set in the instrument (e.g."Geo 18") appears briefly after switch-on or is specified on a label. Table GEO VALUES 3000e shows the geo values for European countries. Table GEO VALUES 6000e/7500e shows the geo values for different gravitation zones. 7.3.
BBA422 / BBA425 / BBK422 Appendix Geographical latitude Geo value Country 47°44' – 49°46' 19 Slovakia 45°26' – 46°35' 18 Slovenia 36°00' – 43°47' 15 Spain 55°20' – 62°00' 24* Sweden 62°00' – 69°04' 26 45°49' – 47°49' 18 Switzerland 35°51' – 42°06' 16 Turkey 49°00' – 55°00' 21* United Kingdom 55°00' – 62°00' 23 * factory setting User manual 22011381A 03/05 41
BBA422 / BBA425 / BBK422 Appendix 7.3.
BBA422 / BBA425 / BBK422 7.4 Appendix Sample protocols Weighing with tare G T N 0.1085 kg 0.0145 kg 0.0940 kg Dynamic weighing Dyn WT T Printout with header 43.52 kg 3.78 kg METTLER TOLEDO www.mt.com G T N G = Gross weight N = Net weight T = Tare 0.1085 kg 0.0145 kg 0.0940 kg Dyn WT = dynamically determined weight Protocol of the scale settings (menu point List, see page 26) SOFTWARE VER 7-0-1.01b SCALE ------------METROLO :NO APPr SNR :0000000 Scale Build SCAL.TYP :2MULT.RN BAS.UNIT :kg SCL.
Index 8 BBA422 / BBA425 / BBK422 Index A Accessories ............................ 38 Adjustment ............................. 19 Alignment................................. 9 Ambient conditions.................. 35 Applications ........................... 34 Average ........................... 13, 21 C Calibrate ................................ 19 Chain tare .............................. 12 Continuous mode.................... 30 D Dimensions ............................ 37 Display ......................
*22011381A* 22011381A Subject to technical changes © Mettler-Toledo (Albstadt) GmbH Mettler-Toledo (Albstadt) GmbH D-72458 Albstadt Tel. ++49-7431-14 0, Fax ++49-7431-14 232 Internet: http://www.mt.