Technical data

METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 3-78
2. Calib Chk Parms
The Calibration Check Parameters sub-block section enables you to specify the test
parameters (default tolerance, number of test weight checks and alarm level) used in the
JAGXTREME calibration test weight check procedure. The user will enter the number of
test points and the starting value of the tolerance permitted between the test weight and
the weight displayed on the scale.
1. Press ENTER at the Calib Chk Parms prompt to configure this sub-block.
2. At the Tol ? prompt, enter the calibration check tolerance in weighing units.
3. At the Test Pts ? prompt, enter the number of test points to be used for the
calibration check (0-99).
4. At the Ann Lev? prompt use the SELECT keys to choose the announcement level
action to be performed when the calibration interval has expired.
Alarm (alarm displayed on lower display)
Log (alarm displayed on lower display and recorded in log file)
Email (alarm displayed on lower display and Email sent)
Disable Sale (alarm displayed on lower display and scale disabled)
5. Press ENTER to continue to the next sub-block or ESCAPE to exit setup.
Ann Lev
Test Pts ?
Calib Chk Parms
Tol ?