Technical data

METTLER TOLEDO JAGXTREME Terminal Technical Manual
(9/00) 7-12
Error Message Description Probable Cause Remedy
OUT_OF_MEMORY The JAGXTREME software cannot
get the dynamic memory it needs
to continue running.
The system is using more
dynamic heap memory
than is available or the heap
memory has become
fragmented.
Reduce the size of a JagBASIC
program. Eliminate unnecessary
spaces in the program. Reduce
the number of variables. Reduce
the size of the arrays.
When chaining JagBASIC
programs, always chain in the
largest program first to reduce
memory fragmentation.
Eliminate unused network
connections, serial connections,
and printer templates.
OUT_OF_ZERO_RANG Operator has attempted to zero
the scale outside of the legal
zeroing range.
The zeroing limits are set up
in the Application Envn, Zero
Operation menu.
Change zeroing range in set up,
if necessary.
Overflow JagBASIC programming error. A JagBASIC program causes
an overflow error by
exceeding certain system
limits. The maximum size of
the gosub stack, the for-
next stack, and the while-
wend stack is 9 entries
each.
Overflow errors can also be
caused by particular
language syntax errors.
Correct JagBASIC program.
PLC_COMM_ERROR There was a timeout in the
communications between the
terminal and a PLC. This error is
only reported after there has been
a successful connection and
communication fails.
The most likely cause is
improper wiring or wiring
termination, or there was a
fault at the PLC. Entering and
exiting setup at the terminal
can also cause this error
because the PLC adapter is
reset.
Check wiring and wiring
terminations. Verify PLC
operation.
Note: The JAGXTREME terminal
will automatically recover
communications after an
intermittent failure.
Prbus config err Attempts to configure the Profibus
card failed.
Most likely, this error is
caused by a Profibus card
hardware failure.
If problem persists, replace the
Profibus card.
PRINT REQUESTED The operator has re- quested a
Demand Print through the Control
Pan el.
None. None.
PRINT_IN_PROGRES The operator has requested a
second demand print while the
first is in progress.
None. None.
PRINT_NOT_READY Scale is in motion while
attempting to print.
None. None.
Profbus init err The Profibus card could not be
initialized.
Bad Profibus card. Replace Profibus card.