Function check
Tap pan support (7) with your hand and check whether the balance display reacts.
Check whether the display shows a constant value when the pan support is at rest.
If this is not the case, the length of spacer bolt (12) must be corrected and the seating of draft
shield (5) checked, see also section 6, “What if...?”
Optimum configuration (the balance)
For optimum functioning of the infrared dryer, the balance must be configured as followed:
LJ16: Stability detector ASd-2-
Integration time Int-2-
Setting the configuration, see section 5.1.
LP16: Stability detector ASd-2-
Pause 0
Vibration adapter
Weighing process adapter
Setting the configuration, see operating instructions of the PM balance used.
Leveling the balance
Turn screw feet (1) so that the air bubble is in the middle of level (2).
Note
After every location change, the balance should be releveled.