Operating instructions

IES Corporation, 2324 SE 11
th
Ave, Portland, OR 97214
PHONE: 1-800-541-0852/1-503-230-0646 FAX: 1-503-235-2535 e-mail: office@iescorp.com
204
TOLEDO
ID SERVICE MODE / Toledo
Although these instruction are written for the ID5 the other ID's are the same except for the displayed message.
FAQ: See Scale Parameters to change units. See External Calibration for internal cal overwrite.
Enter into SERVICE MODE
Switch off terminal
Unplug weighing platform
Switch on terminal. PLUG IN appears in the display
Press key 0until SERVICE MODE appears in the display
Switch off terminal
Attach weighing platform
Switch on the terminal. SERVICE MODE appears briefly in the display followed by the first program block
(RETURN).
______________________________________________________________________________________
Overview of program block structure
RETURN return allows service mode to quit without any changes
RESET reset clears scale number, cell configuration reset to factory setting
NATION nation - country is selected
SCALE PARAMETERS weighing platform, type, maximum capacity, graduations, units
LINEARITY weighing platform can be assigned a new linearity code
CALIBRATION internal calibration or external calibration, internal overwrite
SAVE PARAMETERS the selected configuration is stored
______________________________________________________________________________________
Step through the program blocks by pressing the NO button (T)
Enter a program block by press the YES button ( ⊂→ )
Key Functions: Yes = ⊂→ No = T
RETURN BLOCK
Press YES to return to weighing mode- normal operation
Press NO to enter service mode
RESET BLOCK
Press YES to reset weighing platform, certification capability, maximum capacity, range and graduation to factory
settings. Display shows NO RESET press NO , display shows RESET ALL press YES. Display shows POWER OFF,
Switch off the power then on again, the scale will calibrate internally.
NATION
United States of America = USA
SCALE PARAMETERS
Press YES to enter, and select with YES & NO keys
W+M approval noncertified = NO W+M approval certified = W+M approval
KA15 NO key steps through platforms, YES when desired platform is displayed
CAP maximum capacity of the weighing platform is displayed, set maximum operating
capacity (also can be selected in pounds "lbs" - the unit the capacity is set for will be the unit displayed in the
normal operating mode )
MR MR represents multirange, ( Press Yes/No for resolution setting of multirange )
D D represents single range. ( Press Yes/No for resolution setting of single range )