Operating instructions METTLER TOLEDO AM/PM Balances ; _ 0.0 000 9 ;_ 0.0 ;_ 0.
Thank you very much for the confidence you have shown in our products by choosing a METTLER TOLEDO precision or analytical balance. To obtain complete satisfaction from your balance, it is essential that you read through these operating instructions carefully. These operating instructions apply to balances of the AM and PM series with a weighing range up to 6 kg.
Page 4 5 6 Preparation Choosing the location Mounting the weighing pan / levelling the balance Installing the draft shield Operation Control elements / connections / display METTLER DeltaTrac / METTLER DeltaRange Switching display on and off Simple weighing / taring Single-key (control bar) operation 8 9 10 11 12 Menu Overview Calibrating Adapting to the type of weighing (weighing process adapter) Adapting to the ambient conditions (vibration adapter) 14 15 16 17 Configuring Overview Configuration
Preparation Select the proper location For best results choose a suitable location for your balance. A firm, vibration-free location as horizontal as possible Avoid exposure to direct sunlight No extreme temperature changes 30° 20° 10° 0° No draft Despite a possible unfavorable location, your balance can still produce accurate weighing results: In this case you should adjust the vibration adapter accordingly. For procedures refer to section «Menu».
Mount the weighing pan / platform PM balances with weighing pan Place pan support 11 and weighing pan 10 on tapered pin 17; connect power cable 20. 10 11 20 17 0.0 0 9 10 20 11 40 0.0 0 PM balances with weighing platform Install platform support 11 with the four pins resting on the four rubber grommets 40; then position weighing platform 10 on platform support 11; connect power cable 20. 9 AM balances: see page 7.
Preparation Don't forget to install the draft shield To ensure that the weighing results are not influenced by the drafts (particularly for sensitive measuring ranges) we recommend the installation of a draft shield. PM balances with 1 mg readability Mounting procedures 1 and 2 are needed only if the bottom element 12 has not been installed at the factory. 1 Remove weighing pan 10 and pan support 11. Place bottom element 12 on balance, and rotate to stop. Replace pan support 11 and weighing pan 10.
AM balances Mount glass draft shield 13 on scale rotated by 45°. Note that tapered pin 17 must be visible through bottom hole 27. Now rotate glass draft shield clockwise to stop. 1 Verify that the draft shield is aligned parallel to the balance and is properly engaged in bayonet base. 2 Install ring 25 and weighing pan 10, close sliding door 26. Note: Ring 25 protects the weighing pan from lateral impacts.
Operation Control elements and connections 3 2 4 5 1 8 6 7 8 9 1 2 3 4 5 6 7 8 9 Control bar Program cassette Cover Connector for data interface Connector for METTLER TOLEDO GM instruments Fuse holder (with spare fuse) Power socket Screw feet (level adjustment) Level indicator 31 31a 31b 31c 31d 31e 31f 32 33 34 Status indicator Vibration adapter Weighing process adapter Weight status Automatic zero correction (Autozero) Special status of digital display 1) Stability control Digital display U
The METTLER DeltaTrac This dynamic graphic indicator with 60 radial segments is incorporated in all balance models. METTLER DeltaTrac shows you a graphic representation of the numerical values shown in the digital display. With absolute weighing, subtractive weighing, weighing-in and formula weighing, the dynamic display indicates the weighing range used up and that still remaining.
Operation Switch on display (▼ Display changes automatically) Display switched off (standby) 8.8.8.8.8.8.8 kg oGNt /k%t G 0.00 B N T A Z PCS Stk 9 104000 . .. Briefly press controll bar; all display segments light up briefly (automatic system and display check). Software No. (e.g. 10.40.00, for information only). -.-.-.-.-.-.-.-.-.-.0.0.0 9 Display subsequently indicates zero (weighing mode).
Simple weighing Caution: Before the balance is used for the first time, it must be calibrated (see «Calibration» in section «Menu»). -.-.-.-.0.0.0 9 0.00 0.00 9 9 Weighing mode 11 1 8 2 0 3 9 . .. Load weighing sample 1 250009 .. Wait for stability and read result (Stability is reached as soon as stability detector fades). Taring (▼ display changes automatically) Weighing mode -.-.-.-.0.0.0 9 0.00 0.00 9 Load an empty container or packing. 1 1 2 1.01 . .
Operation Simple operation with a control bar Switching on/off, calibrating, configuring: All of these operations can be initiated with the control bar. You can even use this single control bar to activate the applications incorporated as standard. For more convenient operation of the applications, the GM303 Terminal (direct function, switch and print key) from the accessories can be used. Note these symbols… …You will find them throughout the operating instructions and short-form instructions.
Introducing the symbols Try to familiarize yourself with the key symbols with the aid of the following example. Switch on the display and remove weight from weighing pan. Now try to select and change the weighing process adapter 31b. -.-.-.-.0.0.0 9 Press control bar… 15-cal. . …keep depressed… 31b …release! briefly press control bar! Note: If the display automatically returns to zero (weighing mode) 3 seconds after the control bar was last pressed, simply begin the procedure again.
Menu A clear Menu We distinguish between two levels of software. The first, simpler level, we call the Menu. It can be activated by pressing and holding the control bar. The second software level is called the Configuration file. Details can be found in section «Configuring». When the Menu is activated, you can… 15-cal. . …calibrate your balance, …use the weighing process adapter to set the weighing modes or the weighing sample, and 0.
-cal- Calibrate your balance Before the balance is used for the first time, it must be calibrated (to take the acceleration due to gravity into account). Caution: To obtain accurate results we recomment that you connect the balance to the line 30 min before calibrating (60 min for AM balances). 0.00 -.-.-.-.0.0.0 9 Start calibration with display switched on by pressing and holding the control bar (weighing mode, display zero with empty weighing pan). 9 15-cal. .
Menu Adapt your balance to the type of weighing (Weighing process adapter)… Access -.-.-.-.0.0.0 9 Weighing mode 15-cal. . With this adapter you will optimize the display speed of the last digits as a function of weighing mode: For instance, for fine dispensing of powders even the last place of the digital display must be continuously recognizable. This is not the case for absolute weighing; the weighing process adapter thus suppresses the display during the weighing process.
…as well as to the ambient conditions (Vibration adapter) Access -.-.-.-.0.0.0 9 Weighing mode 15-cal. . Under vibration-free conditions adjust the adapter to obtain results within the shortest possible time. However, if you are operating in an environment with severe vibrations or drafts, adjust the adapter to obtain reliable results, even under adverse conditions.
Configuring 18 Special requirements need special settings within the configuration file Your balance has been factory-set to a standard configuration, i.e. the settings in the configuration file correspond to the most common user requirements. If you wish to change these settings to meet special requirements you must access the configuration file and change the settings according to your needs.
uNI T Unit selection, applications, etc. uNI T1 Selection of basic unit uNI T2 Selection of second unit and applications p¨TT2 Print/transfer command / Starting animal weighing mode Switch on status displays i -face Adapting to peripheral units s . Transmission mode 6 Baud rate p Parity pause Pause between data transfers and handshake Please see the following pages for more information on the individual setting possibilities.
Configuring Configuring – a quick introduction ------- Access Start from standby, i.e. display switched off. Now, press control bar and release only when display shows –Conf.–. Display now automatically changes over to –Reset–. -cONf¨eseT ¨ e s e T s c a l e Selecting sectors The 4 sectors can be selected by briefly pressing the control bar. Note: The –End– display between the sectors –I-Face– and –Reset– indicates the end of the four sectors.
Have you found the setting –Unit 1–? If not, switch off the display by lifting the control bar. Then restart by accessing the configuration file. This time it's sure to work. uNI T1 Change setting, e.g. from –g– (gram) to –ct– (carat): Briefly press control bar several times until carat (ct) appears in display. g Press briefly x times uNI T1 CT uNI T1 CT uNI T2 g p¨ TT2 oFF Auto e ND 0.
Configuring ¨eseT Standard setting and record printout Access Symbols Standby -------cONf- Press control bar briefly Press and hold control bar until required display appears Display changes automatically ¨eseT Important Return to weighing mode by keeping control bar depressed until zero indication appears. If you do not press control bar for 40 sec, the balance will automatically return to weighing mode.
¨eseT Settings NO y e sT 88 Standard setting Yes / No For resetting your balance to standard configuration, select sector –reset–. Now press control bar until –yes– is displayed. By pressing and holding the control bar again until –End– or zero appears, you acknowledge the resetting (weighing mode appears after –End– indication). Your balance is now reset to the original factory setting.
Configuring scal e Setting balance operating parameters Access Standby Symbols Press control bar briefly -------cONf- Press and hold control bar until required display appears ¨eseT Display changes automatically scal e asD -2- Standard setting -3- Selectable setting Important Return to weighing mode, always with control bar depressed till zero indication. If you do not press control bar for 40 sec, the balance will automatically return to weighing mode. Modified settings will be recorded.
Settings scal e as D88 -2-D -3-D -1-D -4-D Automatic stability detection The stability detector 31f lights up when the balance is unstable. At the same time, the data interface is blocked until the weighing result is stable (except for data transfer mode «S» being set to –All– or –Cont–; see sector –I-Face–). -1-D Weighing speed: very fast Reproduction: good -2-D -3-D -4-D D. D88 0.1 0.2 10 0.5 5 1 fast very good Selection of display sequences (Readability) Example for 0.
Configuring uNI Tf- Unit selection, Applications, Print/Transfer command, Status display Access Symbols Standby Press control bar briefly -------cONf- Press and hold control bar until required display appears ¨eseT Display changes automatically scal e uNI T1 uNI Tf- g Standard setting kg - Selectable setting Important Return to weighing mode always with control bar depressed until zero indication.
Setting options uNI Te uNI T1 uNI T2 p¨ T g kg g lb k oz C.M. ozt ct tl dwt GN g x times g PCS % Stk Off ON aUTO ON Basic weighing unit In addition to the weight unit g, the following weight units can be selected for weighing: tl gram kilogram pound ounces troy ounces tael kg g ozt oz lb carat no unit display (display value in g) dwt pennyweight ct C.M.
Configuring i -face Adaptation to external equipment Access Symbols Standby Press control bar briefly ------- Press and hold control bar until required display appears -cONf- Display changes automatically ¨eseT scal e s . I T1 sTB Standard setting all Selectable setting uNI Tfi -face Important Always return to weighing mode by keeping control bar depressed until zero indication appears. If you do not press control bar for 40 sec, the balance will automatically return to weighing mode.
Setting options i -face s . I T1 sTB all cONT aUTO Datenübertragungsmodus sTB all aUTO cONT B . I T1 2400 4800 1200 x times Baud rate The baud rate is the transfer rate unit for serial data transmission in bit/sec: 110 p . I T1 pa u s e aU eND -e- -o- -s- -m- 1 2 h .0 Off ON The next possible stable value is transferred after initiation of print/transfer command (after release by stability detector).
Configuring How to protect the new settings in the configuration file After having completed your configuration, you can protect your settings against inadvertent changes. Proceed as follows: To protect the program cassette from possible interface, it is absolutely essential to first disconnect the line cable 20. 1 Now remove cover 3, for instance with a screwdriver. The program cassette is now visible. 2 Grasp bracket of program cassette 2, and carefully pull out cassette.
In the aperture of the program cassette 2, the shorting jumper 24 is now visible. It can be lifted off from the contact pins (e.g. with a small pencil). 4 not secured To protect your settings in the configuration file from involuntary changes, position the jumper across both contact pins (position secured). secured 5 Reinsert the program cassette 2. Ensure that you push it in firmly right to the stop. Then press on the protective cover 3. Connect power cable 20.
Applications Standard applications at a keystroke Counting in or out of a container, percent formula weighing of powders or liquids, plus/minus checks of fill quantities, weighing in grams or in a second, selectable unit, or even animal weighing: all this and more can be performed with the applications built in as standard. You can choose between the following applications: g ct 175 PCS Animal weighing Animal weighings can also be performed simply and rapidly.
g ct Switching weight unit Configuration Working with two units Standby u N 0. I 00 9 uNI T CT Basic unit, e.g. –g– (gram) ------- till -cONf0.00 ¨eseT scal e 9 In the configuration file, select the setting –Unit 1– in the –Unit– sector: The standard setting has the basic unit gram (g). If you wish to change this, press the control bar repeatedly until the desired unit appears. uNI till uNI Tf- 0.00 uNI T Press briefly x times Basic unit u N I0.
Applications 175 PCS Piece counting (fixed reference number – 10 pieces, 1 item at least 1/4 digit) Configuration Standby Counting into a container ------0.00 9 u N I0.0 0 9 Place empty container onto balance, tare u 6 5 .8 7 9 Add 10 items to container (to obtain reference) -cONf- 10 ¨eseT Select –Unit 2– in sector –Unit– of configuration file 0.00 scal e 9 uNI Tf- till s e T 1.0 PCS u N I 1.
Switch piece count/weight readout Counting out of a container 8.8.8.8.1.7.5 Return to weighing mode uNIT 8.8.8.8.1.7.5 0.00 9 u N I 0.0 0 9 9 - 6 5.8 7 9 g 1.1.5.2.7.3 uNIT PCS 10 g PCS Place filled container on balance, tare 0.00 Select piece counting PCS till 0.00 0.00 Applications s e T 1. 0 PCS u N - 1.0 PCS u - 1 7.
Applications % Plus/minus and percent weighing (tolerance limit ± 2,5%, 100 % = target weight, min. weight = 100 digit) Configuration Set target weight Standby 5 0 0 .0 0 0.00 ------- till -cONf- ¨eseT 0.00 9 9 seT 1 0 0% 9 - 1 0 0.0 0 % Place target weight on pan Press control bar until –SEt 100 %– appears (target weight will be stored) Select –Unit 2– in sector –Unit– of configuration file scal e 0.00 u - 1 0.
Switch percent weighing/weighing mode 9.8.7.0 Coarse weighing-in u N I 0.0 0 % % 0.00 g uNIT 4.9.3.5.0 9 g g u 4 6 7.4 0 % uNIT % % 0.00 9.8.7.
Applications Animal weighing Configuration , Set weighing process adapter 31b to «Animal weighing» - Use vibration adapter 31a to select the desired time cycle (integration time + reading time): Manual • With the external transfer keys (see booklet «Technical data and adcessories») • With the control bar of your balance (set print command to –Prt on–, see «print/transfer command» in this section): 7 5 0 8.3 0 — 3 sec. _ 5 sec. See also section «Operation».
p¨INT Print/transfer command Configuration Initiate print/transfer command Standby 2 5 2 .1 0 ------- 9 Press control bar until –Print– appears in the display. Weighing results are now printed out by an attached printer. Transfer commands may be initiated as well. -cONfp¨INT ¨eseT scal e Select –Prt off– in sector –Unit– of configuration file.
What if… A breakdown should occur anyhow Display Definition Cause Correction Display blank - No Power Scale switched off Power cable disconnected Temporary disturbance - - Incorrect operating voltage Line fuse defective In case of repetition Pan support or/and pan not in place - Zero not defined < - - - 0.0.0 9 <-----> ∂™™™™™ƒ - Off- < - - - 2.
Display
Miscellaneous 42 How to adjust the operating voltage 10 11 20 17 21 22 0.0 0 9 10 20 11 22 21 40 0.0 0 9 Make sure the power cable 20 is disconnected. With the power cable connected, the inside of the balance is live even if the display is blank! Lift off weighing pan 10 and pan support 11; remove screw 21 and carefully lift off upper section of housing 22. Reset voltage selector 23 with a screwdriver. Carefully lower upper section of housing 22 vertically onto the balance.
How to change the protective cover 10 11 10 11 0.0 12/15 16 0 41 16 0.0 0 9 9 A soiled protective cover can be exchanged as follows: Lift off weighing pan 10 and pan holder 11. For balances with round weighing pan, rotate bottom plate 12 or retaining ring 15 until disengaged, lift off. Position balance on its side. Remove control bar 1 by applying pressure toward A. It will disengage and can be lifted off in the direction of B. Remove protective cover 16.
Miscellaneous 44 You can't know all the words Calibrating Adapting the balance to a reference weight LCD Liquid crystal display Configuration cycle A run through the configuration file Menu Configuration file A second level, lockable with the jumper with variable parameters and selectable applications as additions to the menu, see «Configuring» The first level, consisting of calibration, adaptation of weighing process and vibration, can be extended with applications, see « Operation» and «Applica
Technical data for individual models AM50 AM100 PM100 PM200 PM400 PM1200 PM5003 PM480 Comparator DeltaRange PM2500 DeltaRange 0.1 mg 0.1 mg 0.001 g 0.001 g 0.001 g 0.001 g 0.001 g 0.01 g 0.01 g - - - - - - - 0.001 g 0.001 g 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g - - - - - - - 80 g 500 g Taring range (by subtraction) 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g Reproducibility (s) 0.1 mg 0.1 mg 0.5 mg 0.5 mg 0.001 g 0.
Technical data PM300 PM600 PM2000 PM4000 PM6100 PM4800 PM3000 DeltaRange PM6000 PM6 0.01 g 0.01 g 0.01 g 0.01 g 0.01 g 0.1 g 0.1 g 0.1 g 1g - - - - - 0.01 g - - - 310 g 610 g 2100 g 4100 g 6100 g 4100 g 3100 g 6100 g 6100 g - - - - - 800 g - - - Taring range (subtraktiv) 310 g 610 g 2100 g 4100 g 6100 g 4100 g 3100 g 6100 g 6100 g Reproducibility (s) 0.003 g 0.005 g 0.005 g 0.01 g 0.01 g 0.03 g 0.03 g 0.05 g 0.3 g - - - - - 0.
Printed on 100 % chlorine-free Paper, for the sake of our environment.
To protect your METTLER TOLEDO product's future: METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come. Please send for full details about our attractive terms of service. Thank you. P702395 Subject to technical changes and to the availability of the accessories supplied with the instruments. © Mettler-Toledo GmbH 1999 702395C Printed in Switzerland 9910/2.