Operation Manual Multi-parameter Transmitter M400/2(X)H, M400G/2XH Transmitter Multiparameter M400/2(X)H, M400G/2XH 30 031 683
© 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter M400/2(X)H, M400G/2XH 30 031 683
Transmitter M400/2(X)H, M400G/2XH 3 Operation Manual Multi-parameter Transmitter M400/2(X)H, M400G/2XH © 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Transmitter M400/2(X)H, M400G/2XH 30 031 683
Transmitter M400/2(X)H, M400G/2XH © 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland 4 Transmitter M400/2(X)H, M400G/2XH 30 031 683
Transmitter M400/2(X)H, M400G/2XH 5 Content 1 Introduction_ ___________________________________________________________________________________________ 9 2 Safety instructions______________________________________________________________________________________ 10 2.1 Definition of equipment and documentation symbols and designations________________________________________ 10 2.2 Correct disposal of the unit _ ________________________________________________________________________ 11 2.
Transmitter M400/2(X)H, M400G/2XH 7.3 7.4 7.5 7.6 7.7 6 Calibration of amperometric oxygen sensors____________________________________________________________ 38 7.3.1 One-point calibration for amperometric oxygen sensors____________________________________________ 39 7.3.1.1 Auto mode_ ____________________________________________________________________ 39 7.3.1.2 Manual mode___________________________________________________________________ 40 7.3.
Transmitter M400/2(X)H, M400G/2XH 7 10 PID setup_ ____________________________________________________________________________________________ 70 10.1 Enter PID setup___________________________________________________________________________________ 71 10.2 PID auto / manual_________________________________________________________________________________ 71 10.3 Mode___________________________________________________________________________________________ 71 10.3.
Transmitter M400/2(X)H, M400G/2XH 19 8 Buffer tables__________________________________________________________________________________________ 105 19.1 Standard pH buffers_ _____________________________________________________________________________ 105 19.1.1 Mettler-9________________________________________________________________________________ 105 19.1.2 Mettler-10_ _____________________________________________________________________________ 106 19.1.
Transmitter M400/2(X)H, M400G/2XH 1 9 Introduction Statement of Intended Use – The 2-wire M400 multi-parameter transmitter is a single- channel online process instrument with HART® communication capabilities for measuring various properties of fluids and gases. These include Conductivity, Dissolved Oxygen, and pH / ORP. The M400 is available in two different levels. The level indicates the amount of measurement parameters which can be covered.
Transmitter M400/2(X)H, M400G/2XH 2 10 Safety instructions This manual includes safety information with the following designations and formats. 2.1 a a h a Definition of equipment and documentation symbols and designations WARNING: POTENTIAL FOR PERSONAL INJURY. CAUTION: possible instrument damage or malfunction. NOTE: Important operating information.
Transmitter M400/2(X)H, M400G/2XH 11 h NOTE: PROCESS UPSETS Because process and safety conditions may depend on consistent operation of this transmitter, provide appropriate means to maintain operation during sensor cleaning, replacement or sensor or instrument calibration. h NOTE: This is a 2-wire-product with two active 4–20 mA analog output. 2.2 Correct disposal of the unit When the transmitter is finally removed from service, observe all local environmental regulations for proper disposal.
Transmitter M400/2(X)H, M400G/2XH 2.3 12 Ex Instructions for M400 series multi-parameter transmitters M400 series multi-parameter transmitters are produced by Mettler-Toledo AG. It has passed the inspection of IECEx and conforms to following standards: – IEC 60079-0 : 2011 Edition: 6.0 Explosive atmospheres – Part 0: General requirements – IEC 60079-11 : 2011 Edition: 6.
Transmitter M400/2(X)H, M400G/2XH 13 Terminal 10, 11 12, 13 1, 2; 3, 4; 6, 7; 8, 9; P,Q Function Aout1 Aout2 Digital Input OC Output Analog Input Safety Parameters Ui = 30 V Ii = 100 mA Ui = 30 V Ii = 100 mA Ui = 30 V Ii = 100 mA Ui = 30 V Ii = 100 mA Ui = 30 V Ii = 100 mA Pi = 0.8 W Pi = 0.8 W Pi = 0.8 W Pi = 0.8 W Pi = 0.8 W Li ≈ 0 Li ≈ 0 Li ≈ 0 Li ≈ 0 Li ≈ 0 Ci = 15 nF Ci = 15 nF Ci ≈ 0 Ci ≈ 0 Ci = 15 nF N, O RS485 Sensor Ui = 30 V Uo = 5.88 V Ii = 100 mA Io = 54 mA Pi = 0.
Transmitter M400/2(X)H, M400G/2XH 3 14 Unit overview The M400 models are available in 1/2DIN case size. The M400 models provide an integral IP66/NEMA4X housing for wall- or pipe mount. 3.1 Overview 1/2DIN 5.90 [150] 1 M300 3 5.
Transmitter M400/2(X)H, M400G/2XH 15 3.2 Control / Navigation Keys 3.2.
Transmitter M400/2(X)H, M400G/2XH 3.2.2.2 16 Escape Press the and c key simultaneously (escape) to return to the Measurement mode. 3.2.2.3 ENTER Use the e key to confirm action or selections. 3.2.2.4 Menu Press the key to access the main Menu. 3.2.2.5 Calibration mode Press the c key to enter Calibration mode. 3.2.2.6 Info mode Press the . key to enter Info mode. 3.2.
Transmitter M400/2(X)H, M400G/2XH 3.2.5 17 Navigation with u in Display If a u is displayed on the bottom right hand corner of the display, you can use the c or the key to navigate to it. If you click [ENTER] you will navigate backwards through the menu (go back one screen). This can be a very useful option to move back up the menu tree without having to exit into the measuring mode and re-enter the menu. 3.2.
Transmitter M400/2(X)H, M400G/2XH 18 4 Installation instruction 4.1 Unpacking and inspection of equipment Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Do not discard the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Mettler-Toledo immediately. 4.1.
Transmitter M400/2(X)H, M400G/2XH 4.1.2 19 Installation procedure General: – Orient the transmitter so that the cable grips face downward. – Wiring routed through the cable grips shall be suitable for use in wet locations. – In order provide IP66 enclosure ratings, all cable glands must be in place. Each cable gland must be filled using a cable, or suitable Cable Gland Hole Seal. For Wall Mount: – Remove rear cover from front housing.
Transmitter M400/2(X)H, M400G/2XH 4.1.4 20 1/2DIN version – Dimension drawings 114 mm/4.49" 27 mm 1.06" 150 mm / 5.9" 136 mm / 5.35" 150 mm/5.9" 28 mm 1.10" 43 mm 1.69" 90 mm 3.54" 90 mm / 3.54" 75 mm / 2.95" 6 mm/ 0.236” 90 mm / 3.54" 137 mm (+ 0,5 mm/– 0,0 mm) 5.39" (+ 0.02"/– 0.00") 4.1.5 137 mm (+ 0,5 mm/– 0,0 mm) 5.39" (+ 0.02"/– 0.00") 80 mm / 3.15" 35mm/ 1.38” 1/2DIN version – Pipe mounting 40 ... 60 mm 1.57... 2.
Transmitter M400/2(X)H, M400G/2XH 4.2 21 Connection of power supply All connections to the transmitter are made on the rear panel of all models. a Be sure power to all wires is turned off before proceeding with the installation. A two-terminal connector on the rear panel of all M400 models is provided for power connection. All M400 models are designed to operate from a 14–30 VDC power source.
Transmitter M400/2(X)H, M400G/2XH 22 4.3 Connector PIN definition 4.3.1 Terminal Block (TB) Definitions Power connections are labeled – N for Neutral and + L for Line, for 14–30 VDC. A 1 Q 9 TB2 10 15 TB1 Power connections are labeled A01+ / HART and A01– / HART resp. A02+ and A02– for 14 to 30 VDC. TB1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 DI1+ DI1– DI2+ DI2– Not used OC1+ OC1– OC2+ OC2– AO1+/HART AO1–/HART AO2+ AO2– Not used OC1, 2 = Open collector output 1, 2.
Transmitter M400/2(X)H, M400G/2XH 4.3.
Transmitter M400/2(X)H, M400G/2XH 4.3.
Transmitter M400/2(X)H, M400G/2XH InPro 3250i/SG/120 Art.-Nr. 52 005 373 25 4.4 Connection of ISM (digital) sensors 4.4.1 Connection of ISM sensors for pH/ORP, Cond 4-e and amperometric oxygen measurement * ** h NOTE: Connect the sensor and screw the plug head clockwise (hand tight). 4.4.
Transmitter M400/2(X)H, M400G/2XH h 26 4.5 Connection of analog sensors 4.5.1 Connection of analog sensor for pH / ORP NOTE: Cable lengths > 20 m can worsen the response during pH measurement. Be sure to observe the sensor instruction manual.
Transmitter M400/2(X)H, M400G/2XH 27 4.5.2 TB2 – Typical wiring for analog pH/ORP sensor 4.5.2.1 Example 1 pH measurement without Solution Ground M L K J I H G Jumper F E B transparent red Cable white green yellowgreen A Temperature probe h Combination pH electrode NOTE: Jumber terminals G and F Wire Colors only valid for connection with VP caple; blue and grey not connected.
Transmitter M400/2(X)H, M400G/2XH 4.5.2.2 28 Example 2 pH measurement with Solution Ground M L K J I H G F E B A Temperature probe h transparent red blue white green yellowgreen Cable Combination pH electrode with RTD and SG NOTE: Wire colors only valid for connection with VP cable, grey not connected.
Transmitter M400/2(X)H, M400G/2XH 4.5.2.
Transmitter M400/2(X)H, M400G/2XH 4.5.2.4 30 Example 4 ORP measurement with pH solution ground electrode (e.g. InPro 3250, InPro 4800 SG).
Transmitter M400/2(X)H, M400G/2XH 4.5.3 h 31 Connection of analog sensor for amperometric oxygen measurement NOTE: Be sure to observe the sensor instruction manual.
Transmitter M400/2(X)H, M400G/2XH 4.5.4 32 TB2 – Typical wiring for analog sensor for Amperometric oxygen measurement M K I H G F E D C B h white green grey blue transparent yellow/green red A NOTE: Wire colors only valid for connection with VP cable, but not connected.
Transmitter M400/2(X)H, M400G/2XH a 33 5 Placing transmitter in, or out, of service 5.1 Placing transmitter in service After connecting the transmitter to power supply circuit, it will be active as soon as the circuit is powered. 5.2 Placing transmitter out of service First disconnect the unit from the main power source, then disconnect all remaining electrical connections. Remove the unit from the wall / panel.
Transmitter M400/2(X)H, M400G/2XH 6 34 Quick Setup (PATH: Menu / Quick Setup) Select Quick Setup and press the [ENTER] key. Enter the security code if necessary (see section 9.2 “Passwords”) h NOTE: Please find the complete description of the Quick Setup routine described in the separate booklet “Quick Setup Guide for Transmitter M400” enclosed in the box. h NOTE: Please do not use Quick Setup menu after configuration of the transmitter, because some of the parameters i.e.
Transmitter M400/2(X)H, M400G/2XH 7 35 Sensor Calibration (PATH: Cal) The calibration key c allows the user one-touch access to sensor calibration and verification features. h NOTE: During Calibration on Channel A, a flashing ”H” (Hold) in the upper left corner of the Display indicates a calibration is in process with a Hold condition active. (The hold output needs to be activated.) See also chapter 3.2.8 “Display”. 7.1 Enter Calibration Mode While in Measurement mode press the c key.
Transmitter M400/2(X)H, M400G/2XH 7.2 36 Conductivity calibration for two- or four-electrode sensors This feature provides the ability to perform a one-point, two-point or process Conductivity resp. Resistivity “Sensor” calibration for two- or four-electrode sensors. The procedure described below works for both types of calibrations. There is no reason to perform a two-point calibration on a two-electrode conductivity sensor.
Transmitter M400/2(X)H, M400G/2XH 37 Enter the value for Point 1 including a decimal point and units. The value in the second text line is the value being measured by the transmitter and sensor in the units selected by the user. Press [ENTER] when this value is stable to perform the calibration. After the calibration the cell multiplier or slope calibration factor ”M” i.e. cell constant and the Adder or offset calibration factor ”A” are displayed.
Transmitter M400/2(X)H, M400G/2XH 7.2.3 38 Process Calibration (Display reflects typical Conductivity sensor calibration) Enter Conductivity Sensor Calibration mode as described in section 7.1 “Enter Calibration Mode” and choose one of the compensation modes (see section 7.2 “Conductivity calibration for twoor four electrode sensors”). Select Process Calibration and press [ENTER]. With conductivity sensors a process calibration is always performed as a slope calibration.
Transmitter M400/2(X)H, M400G/2XH 7.3.1 39 One-point calibration for amperometric oxygen sensors Enter Oxygen calibration mode as described in section 7.1 “Enter Calibration Mode”. A one-point calibration of oxygen sensors is always either a one point slope (i.e. with air) or a zero (offset) calibration. A one point slope calibration is done in air and a one point offset calibration is done at 0 ppb oxygen.
Transmitter M400/2(X)H, M400G/2XH 7.3.1.2 40 Manual mode Enter the value for Point 1 including a decimal point and units. The value in the second text line is the value being measured by the transmitter and sensor in the units selected by the user. Press [ENTER] when this value is stable to perform the calibration. After the calibration the slope “S” and the offset value ”Z” are displayed.
Transmitter M400/2(X)H, M400G/2XH 41 After the calibration the slope ”S” and the offset value ”Z” are displayed. In case of a successful calibration, the calibration values are stored in the cal history* and taken over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort). * only available with ISM sensor. The values will be stored in the sensor. If ”Adjust” or ”Calibrate” are chosen, the message ”Calibration successful” is displayed.
Transmitter M400/2(X)H, M400G/2XH 42 As soon as the stabilisation criteria have been fulfilled the display changes. The display shows now the slope calibration factor S and the offset calibration factor Z. In case of a successful calibration, the calibration values are stored in the cal history* and taken over (Adjust), stored in the cal history* and not taken over (Calibrate) or discarded (Abort). * only available with ISM sensor. The values will be stored in the sensor.
Transmitter M400/2(X)H, M400G/2XH 43 The display shows the buffer the transmitter has recognized (Point 1) and the measured value. As soon as the stabilisation criteria have been fulfilled stabilisation criteria have been fulfilled, the display changes and prompts you to place the electrode in the second buffer. Place the electrode in the second buffer solution and press the [ENTER] key to go on with the calibration.
Transmitter M400/2(X)H, M400G/2XH 7.4.3 44 Process calibration Enter pH calibration mode as described in section 7.1 “Enter Calibration Mode”. Select Process calibration. With pH sensors a process calibration is always performed as a offset calibration. Take a sample and press the [ENTER] key again to store the current measuring Value. To show the ongoing calibration process, A or B (depending on the channel) is blinking in the display.
Transmitter M400/2(X)H, M400G/2XH 45 The user can now enter Point 1. The offset calibration factor is calculated by using the value of Point1 instead of the measured value (line 4, mV = ....) and displayed on the next screen. Z is the newly calculated offset calibration factor. The slope calibration factor S is always 1 and does not enter the calculation. After a successful calibration, the calibration values are taken over (Adjust) or discarded (Calibrate) or (Abort).
Transmitter M400/2(X)H, M400G/2XH 7.5 46 Sensor temperature calibration (only for analog sensors) Enter Sensor calibration mode as described in section 7.1 “Enter Calibration Mode” and select Temperature. 7.5.1 One-Point sensor temperature calibration Select 1 point calibration. Slope or Offset can be selected with the 1 Point calibration. Select Slope to recalculate the Slope factor M (Multiplier) or Offset to recalculate the offset calibration factor A (Adder).
Transmitter M400/2(X)H, M400G/2XH 47 After a successful calibration, the calibration values are taken over (Adjust) or aborted (Calibrate, Abort). If “Adjust” has been chosen, the message “Calibration successful” is displayed. In any case you will get the message “Re-install sensor” and “Press ENTER” on the display. After pressing “ENTER” the M400 returns to the measuring mode. 7.6 Edit sensor calibration constants (only for analog sensor) Enter Calibration mode as described in section 7.
Transmitter M400/2(X)H, M400G/2XH 8 48 Configuration (PATH: Menu / Configure) Configure Measurement Analog Outputs Set Points Alarm/Clean ISM Setup* Display Hold Outputs * Only available in combination with ISM sensors 8.1 Enter configuration mode While in Measurement mode, press the key. Press the or . key to navigate to the Configure – menu and press [ENTER]. 8.2 Measurement (PATH: Menu / Configure / Measurement) Enter configuration mode as described in Section 8.
Transmitter M400/2(X)H, M400G/2XH 8.2.1.1 49 Analog sensor Select sensor type Analog and press [ENTER].
Transmitter M400/2(X)H, M400G/2XH 8.2.2 50 Temperature source (only for analog sensors) (PATH: Menu/Configure/Measurement/Temperature Source) Enter Measurement as described in chapter 8.2 “Measurement”. Select Temperature Source by using the or . key and press [ENTER]. The following options can be chosen: Auto: The transmitter automatically recognizes the temperature source. Use NTC22K: Input will be taken from the sensor attached.
Transmitter M400/2(X)H, M400G/2XH 8.2.3.1 51 Conductivity temperature compensation If during the channel setup (see chapter 8.2.1 “Channel setup”) the parameter conductivity has been chosen or an four-electrode conductivity sensor based on ISM technology is connected to the transmitter, the temperature compensation mode can be selected. Temperature compensation should be matched to the characteristics of the application.
Transmitter M400/2(X)H, M400G/2XH 8.2.3.2 52 Concentration table If during the channel setup (see chapter 8.2.1 “Channel setup”) the parameter conductivity has been chosen or an four-electrode conductivity sensor based on ISM technology is connected to the transmitter, a concentration table can be defined. To specify customers-specific solutions, up to 9 concentration values can be edited in a matrix together with up to 9 temperatures.
Transmitter M400/2(X)H, M400G/2XH h 53 NOTE: The conductivity values at the different temperatures have to increase or decrease from Concentration1 to Concentration2 to Concentration3 etc.. Maxima and/or minima are not permitted. If the conductivity values at Temp1 are increasing with the different concentrations, they have to increase also at the other temperatures.
Transmitter M400/2(X)H, M400G/2XH 54 The units for the slope and the zero point, that will be shown on the display can be chosen. The default setting for the unit of the slope is [%] and can be changed to [pH/mV]. For the zero point the default setting of the unit is [pH] and can be changed to [mV]. Use the c key to move to the input field and select the unit by using the or key. Pressing [ENTER] again will bring up the Save Changes dialog.
Transmitter M400/2(X)H, M400G/2XH 55 If an ISM sensor has been connected resp. configured there is furthermore the option to adjust the polarization voltage for the sensor. Different value can be entered for the measuring mode (Umeaspol) and for the calibration mode (Ucalpol). For entered values 0 mV to –550 mV the connected sensor will be set to a polarization voltage of –500mV. If the entered value is less then –550mV, the connected sensor will set to a polarization voltage of –674mV.
Transmitter M400/2(X)H, M400G/2XH 8.3 56 Analog outputs (PATH: Menu / Configure / Analog Outputs) Enter Configuration mode as described in Section 8.1. “Enter Configuration Mode” and navigate to the menu “Analog Outputs” by using the or . key. Press the [ENTER] key to select this menu, which lets you configure the 4 analog outputs. Once analog outputs have been selected, use the and c buttons to buttons to navigate b etween configurable parameters.
Transmitter M400/2(X)H, M400G/2XH 57 Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made. 8.4 Set points (PATH: Menu / Configure / Set Points) Enter Configuration mode as described in Section 8.1. “Enter Configuration Mode” and navigate to the menu “Set Points” by using the or . key. Press the [ENTER] key to select this menu.
Transmitter M400/2(X)H, M400G/2XH 58 Hold Enter the OC Hold Status of ”Last”, ”On” or ”Off”. This is the state the OC will go to during a hold status. State OC contacts are in normal state until the associated setpoint is exceeded, then the OC is activated and the contact states change. Select ”Inverted” to reverse the normal operating state of the OC (i.e. normally high voltage state is in a low voltage state until the setpoint is exceeded). ”Inverted” OC operation is functional vice versa.
Transmitter M400/2(X)H, M400G/2XH 59 The conditions for alarms are: 1. There is a power failure or power cycling 2. The software watchdog performs a reset 3. Rg is out of tolerance – for example, broken measuring electrode (only for pH; pH/pNa Rg diagnostics detect both pH and pNa membrane glasses) 4. Rr is out of tolerance – for example, coated or depleted reference electrode (only for pH sensors; except pH/pNa) 5.
Transmitter M400/2(X)H, M400G/2XH 60 Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard the entered values and return to the measurement display screen, selecting Yes will save changes made. 8.6 ISM set up (available for pH and oxygen ISM sensors) (PATH: Menu / Configure / ISM Setup) Enter Configuration mode as described in Section 8.1. “Enter Configuration Mode” and navigate to the menu “ISM set up” by using the or . key. Press [ENTER] 8.6.
Transmitter M400/2(X)H, M400G/2XH 61 Time to Maintenance: This timer estimates when the next cleaning cycle should be performed to keep the best possible measurement performance. The timer is influenced by significant changes on the DLI parameters. Time to Maintenance Alarm YES / NO YES / NO R# choose OC The time to maintenance can be reset to the initial value by the menu ”Reset ISM Counter Timer” (see chapter 8.6.5 “Reset ISM counter/timer”).
Transmitter M400/2(X)H, M400G/2XH 62 CIP characteristics: CIP Cycles will be automatically recognized by the sensor. Since CIP cycles will vary in intensity (duration and temperature) for each application the algorithm of the counter recognizes an increase of the measurement temperature above a adjustable limit (parameter Temp in °C).
Transmitter M400/2(X)H, M400G/2XH 63 The Autoclaving Cycle Limit counts the number of autoclaving cycles. If the limit (user defined) is reached, an alarm can be indicated and set to a certain output OC. The following options are possible: Autoclave Max 000 Alarm YES / NO R# choose OC If the Max setting is on 000, the counter functionality is turned off. The alarm will be reset after exchanging the sensor.
Transmitter M400/2(X)H, M400G/2XH 64 Navigate to the menu “DLI Stress Adjustment” by using the and . keys and press [ENTER]. Adjust the Process Stress parameter based on the particular application and/or requirements Low: DLI, TTM and ACT will be increased approximately 25% compared with “Medium”. Medium: Default value, (equal DLI, TTM and ACT values based on former firmware versions of the transmitter). High: DLI, TTM and ACT will be reduced approximately 25% compared with “Medium”.
Transmitter M400/2(X)H, M400G/2XH 8.7.2 65 Resolution This menu allows the setting of the resolution of each displayed value. The accuracy of the measurement is not effected by this setting. Possible settings are 1, 0.1, 0.01, 0.001 or Auto. Pressing the [ENTER] key will bring up the Save Changes dialog. 8.7.3 Backlight This Menu allows the setting of the back light options of the display. Possible settings are On, On 50% or Auto Off 50%.
Transmitter M400/2(X)H, M400G/2XH 8.7.5 66 ISM sensor monitoring (available when ISM sensor connected) The sensor monitoring allows you to display the sensor monitoring details on line 3 and 4 in the display. The following options are possible: Line 3 Off / Time Indicator / Time to Maint / Adapt Cal Timer Line 4 Off / Time Indicator / Time to Maint / Adapt Cal Timer 8.8 Hold analog outputs (PATH: Menu / Configure / Hold Outputs) Enter configuration mode as described in Section 8.
Transmitter M400/2(X)H, M400G/2XH 9 67 System (PATH: Menu / System) System Set Language Passwords Set/Clear Lockout Reset Set Date & Time While in measurement mode press the key. Press the . or key to navigate to ”System” – Menu and press [ENTER]. 9.1 Set Language (PATH: Menu / System / Set Language) This menu allows the configuration of the display language. The following selections are possible: English, French, German, Italian, Spanish, Portuguese, Russian or Japanese (Katakana).
Transmitter M400/2(X)H, M400G/2XH 9.2.1 68 Changing passwords See Section 9.3 on how to enter the passwords menu. Select Change Administrator or Change Operator and set the new password. Press the [ENTER] key and confirm the new password. Press [ENTER] again to bring up the Save Changed dialog. 9.2.2 Configuring menu access for operator See 9.3 on how to enter the passwords Menu. Select Configure Operator to configure the access list for the operator.
Transmitter M400/2(X)H, M400G/2XH 9.4.1 69 Reset system This menu allows the reset of the meter to the factory default settings (setpoints off, analog outputs off, etc.). The meter calibration and the analog output calibration are not affected. Pressing the [ENTER] key after the selection will bring up a confirmation screen. Selecting No will return the user to the measurement mode with no changes. Selecting Yes will reset the meter. 9.4.
Transmitter M400/2(X)H, M400G/2XH 10 70 PID setup (PATH: Menu / PID Setup) PID Setup PID A/M Tune Parameters Mode PID Display Setup PID control is proportional, integral and derivative control action that can provide smooth regulation of a process. Before configuring the transmitter, the following process characteristics must be identified.
Transmitter M400/2(X)H, M400G/2XH 71 Determine the appropriate settings for each of these control parameters based on the shape of the pH process titration curve. Controller with Corner Points Proportional limit +100% value 120% 100% 80% YP Error (%) 60% Direct corner point (value, %) 40% Set point value 20% Deadband + value 0% Deadband - value –20% Reverse corner point (value, %) –40% –60% –80% –100% Proportional limit –100% value –120% 0 2 4 6 8 10 12 14 Process Variable 10.
Transmitter M400/2(X)H, M400G/2XH 10.3.1 72 PID mode This menu assigns a OC or analog output for PID control action as well as details of their operation. Based on the control device being used, select one of the following three paragraphs for use with solenoid valve, pulse input metering pump or analog control. Pulse Length – If using a solenoid valve, select ”OC” and ”PL”, pulse length.
Transmitter M400/2(X)H, M400G/2XH 10.4 73 Tune parameters (PATH: MENU / PID Setup / Tune Parameters) This menu assigns control to a measurement and sets the setpoint, tuning parameters and nonlinear functions of the controller through a series of screens. 10.4.1 PID assignment & tuning Assign the measurement, a, b, c, or d to be controlled after ”PID on_”. Set the Gain (unitless), integral or reset time Tr (minutes) and rate or derivative time Td (minutes) needed for control. Press [ENTER].
Transmitter M400/2(X)H, M400G/2XH 10.5 74 PID display (PATH: Menu/PID Setup/PID Display Setup) This screen enables display of PID control status in the normal measurement mode. When PID Display is selected, the status (Man or Auto) and control output (%) will be displayed on the bottom line. If controlling pH, the reagent will also be displayed. In addition, for the display to be enabled, a measurement must be assigned under Tune Parameters and a OC or analog output must be assigned under Mode.
Transmitter M400/2(X)H, M400G/2XH 11 75 Service (PATH: Menu / Service) Service Tech Service Calibrate Diagnostics While in measurement mode press the key. Press the or . key to navigate to the ”Service” menu and press [ENTER]. The available system configuration options are detailed below. 11.
Transmitter M400/2(X)H, M400G/2XH 11.1.3 76 Display All pixels of the display will be lit for 15 seconds to allow troubleshooting of the display. After 15 seconds the transmitter will return to the normal measuring mode or press [ENTER] to exit sooner. 11.1.4 Keypad For keypad diagnostics, the display will indicate which key is pressed. Pressing [ENTER] will return the transmitter to the normal measuring mode. 11.1.
Transmitter M400/2(X)H, M400G/2XH 11.1.7 77 Read OC The Read OC diagnostic menu shows the state of each OC as defined below. To display OC 5 and 6, press [ENTER]. Press [ENTER] again to exit from this display. 0 = Normal 1 = Inverted. 11.1.8 Set analog outputs This menu enables the user to set all analog outputs to any mA value within the 0–22 mA range. Press [ENTER] to exit from this display. 11.1.9 Read analog outputs This menu shows the mA value of the analog outputs.
Transmitter M400/2(X)H, M400G/2XH 11.2.1 78 Calibrate meter (only for channel A) The M400 transmitter is factory calibrated within specifications. It is not normally necessary to perform meter re-calibration unless extreme conditions cause an out of spec operation shown by Calibration Verification. Periodic verification / re-calibration may also be necessary to meet Q.A. requirements.
Transmitter M400/2(X)H, M400G/2XH 79 Enter the value for Point 1, in milliamps, of the current source connected to the input. The second display line will show the measured current. Press [ENTER] to begin the calibration process. Enter the value for Point 2, in milliamps, of the current source connected to the input. The second display line shows the measured current. Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen.
Transmitter M400/2(X)H, M400G/2XH 80 Enter the value for Point 1 of the calibration according to the resistor connected across the pH glass electrode measuring input. Press [ENTER] to begin the calibration process. Enter the value for Point 2 of the calibration according to the resistor connected across the pH glass electrode measuring input. Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen.
Transmitter M400/2(X)H, M400G/2XH 81 Connect an accurate milliamp meter to the analog output terminals and then adjust the five digit number in the display until the milliamp meter reads 4.00 mA and repeat for 20.00 mA. As the five digit number is increased the output current increases and as the number is decreased the output current decreases.
Transmitter M400/2(X)H, M400G/2XH 12 82 Info (PATH: Info) Info Calibration Data Messages Model/Software Revision ISM Sensor Info* ISM Diagnostics* * Only available in combination with ISM sensors Pressing the . key will display the Info menu with the options Messages, Calibration Data and Model/Software Revision. 12.1 Messages (PATH: Info/Messages) The most recent message is displayed. The up and down arrow keys allow scrolling through the last four messages that have occurred.
Transmitter M400/2(X)H, M400G/2XH 12.3 83 Model / Software revision (PATH: Info/Model/Software Revision) Selecting Model/Software Revision will display the part number, model and the serial number of the transmitter. By using the .
Transmitter M400/2(X)H, M400G/2XH 84 Cal History The calibration history is stored with a time stamp in the ISM sensor and is displayed on the transmitter. The calibration history offers the following information: Fact (Factory calibration): This is the original dataset, determined in the factory. This dataset remains stored in the sensor for reference and cannot be overwritten. Act (Actual adjustment): This is the actual calibration dataset which is used for the measurement.
Transmitter M400/2(X)H, M400G/2XH 85 Time to Maintenance: This timer shows a Time to Maintenance, when the next cleaning cycle should be performed to keep the best possible measurement performance. The Time to Maintenance is indicated in days (d) and percentage (%). For a description of the Time to Maintenance, please see section 8.6 “ISM Setup”. For oxygen sensors, the Time to Maintenance indicates a maintenance cycle for the membrane and electrolyte. Press [ENTER] to exit from this display. Max.
Transmitter M400/2(X)H, M400G/2XH 13 Maintenance 13.1 Front panel cleaning 86 Clean the front panel with a damp soft cloth (water only, no solvents). Gently wipe the surface and dry with a soft cloth.
Transmitter M400/2(X)H, M400G/2XH 14 87 Troubleshooting If the equipment is used in a manner not specified by Mettler-Toledo the protection provided by the equipment may be impaired. Review the table below for possible causes of common problems: Problem Possible Cause – No power to M400. – LCD display contrast set incorrectly. – Hardware failure. Display is blank. Incorrect measurement readings. Measurement readings not stable. – Sensor improperly installed. – Incorrect units multiplier entered.
Transmitter M400/2(X)H, M400G/2XH 14.2 88 Cond (resistive) Error messages / Warning- and Alarm list for ISM sensors Alarms Watchdog time-out* Description SW/System fault Dry Cond sensor* Cell running dry (no measurement solution) Cell deviation* Multiplier out of tolerance** (depends on sensor model). * According to the parameterization of the transmitter (see chapter 8.5.1 “Alarm”; PATH: Menu/Configure/Alarm/Clean/Setup Alarm) ** For further information refer to the sensor documentation 14.
Transmitter M400/2(X)H, M400G/2XH 14.3.2 89 Dual membrane pH electrodes (pH / pNa) Warnings Warning pH slope >102% Warning pH Slope < 90% Warning pH Zero > 8.0 pH Warning pH Zero < 6.0 pH Warning pHGls change < 0.3* Warning pHGls change > 3* Warning pNaGls change< 0. 3* Warning pNaGls change > 3* Description Slope too big Slope too small Zero offset too big Zero offset too small Glass electrode resistance changed by more than factor 0.
Transmitter M400/2(X)H, M400G/2XH 90 14.4 Amperometric O2 Error messages / Warning- and Alarm list 14.4.1 High level oxygen sensors Warnings Warning O2 Slope < – 90 nA Warning O2 Slope > – 35 nA Warning O2 ZeroPt > 0.3 nA Warning O2 ZeroPt < – 0.3 nA Description Slope too big Slope too small Zero offset too big Zero offset too small Alarms Watchdog time-out* Error O2 Slope < –110 nA Error O2 Slope > – 30 nA Error O2 ZeroPt > 0.6 nA Error O2 ZeroPt < – 0.
Transmitter M400/2(X)H, M400G/2XH 14.4.3 91 Trace oxygen sensors Warnings Warning O2 Slope < – 5000 nA Warning O2 Slope > – 3000 nA Warning O2 ZeroPt > 0.5 nA Warning O2 ZeroPt < – 0.5 nA Description Slope too big Slope too small Zero offset too big Zero offset too small Alarms Watchdog time-out Error O2 Slope < – 6000 nA Error O2 Slope > – 2000 nA Error O2 ZeroPt > 1.0 nA Error O2 ZeroPt < – 1.
Transmitter M400/2(X)H, M400G/2XH 92 14.5 Warning- and Alarm indication on the display 14.5.1 Warning indication If there are conditions, which generate a warning, the message will be recorded and can be selected through the menu Messages (PATH: Info / Messages; see also chapter 12.1 “Messages”). According to the configuration of the transmitter the hint ”Failure – Press ENTER” will be shown at line 4 of the display, if a warning or alarm has occurred (see also chapter 8.
Transmitter M400/2(X)H, M400G/2XH 15 93 Accessories and Spare Parts Please contact your local Mettler-Toledo sales office or representative for details for Additional accessories and spare parts. Description Order no.
Transmitter M400/2(X)H, M400G/2XH 94 16 Specifications 16.1 General specifications Conductivity / resistive Specifications Range 0.01 cm–1 constant sensor 0.002 to 200 uS/cm (5000 Ωxcm to 500 MΩ xcm) Range 0.1 cm-1 constant sensor 0.02 to 2000 uS/cm (500 Ω x cm to 50 MΩ x cm) Range lOcnr1 constant sensor 10 to 40,000 uS/cm (25 Ω x cm to 100 KΩ x cm) Display range for 2-e sensor 0 to 40,000 mS/cm (25 Ω x cm to 100 MΩ x cm) Display range for 4-e sensor 0.01 to 650 mS/cm (1.54 Ω x cm to 0.
Transmitter M400/2(X)H, M400G/2XH pH Specifications: pH range Sensor maximum distance pH resolution pH accuracy** mV range mV resolution mV accuracy Temperature input* Temperature measuring range Temperature resolution Temperature accuracy** Temperature repeatability** 95 –2.00 to 16.00 pH Analogue: 10 to 20 m (33 to 65 ft) depending on sensor ISM: 80 m (260 ft) auto / 0.01 / 0.1 / 1 (can be selected) ± 0.02 pH –1500 to 1500 mV auto / 0.01 / 0.
Transmitter M400/2(X)H, M400G/2XH 96 Available Buffer Sets: Standard buffers MT-9 buffers, MT-10 buffers, NIST Technical Buffers, NIST Standard Buffers (DIN 19266:2000–01), JIS Z 8802 buffers, Hach buffers, CIBA (94) buffers, Merck Titrisols-Reidel Fixanals, WTW buffers Dual membrane electrodes pH buffers (pH/pNa) Mettler-pH/pNa buffers (Na+ 3.
Transmitter M400/2(X)H, M400G/2XH 16.2 Electrical specifications Number of current outputs Current outputs Measurement error through analog outputs Analog output configuration Display Running capacity Keypad Languages PID process controller Cycle time Hold input / Alarm contact Connection terminals Digital outputs Digital input Analog input Alarm delay 16.
Transmitter M400/2(X)H, M400G/2XH 16.
Transmitter M400/2(X)H, M400G/2XH 17 99 Default table Common Parameter Alarm Clean Hold outputs DigitalIn Lockout ISM monitor Sub parameter OC delay hysterseis state Power failure Software tailure ChB disconnected OC Hold mode interval Clean time delay hysteresis Lifetime indicator Time to maint Adapt Cal timer CIP cycle counter SIP cycle counter Autoclave cycle counter OC language Passwords All OCs All analog out administrator operator delay Value 2 1 0 inverted No No Yes 1 Hold 0 0 0 0 Yes
Transmitter M400/2(X)H, M400G/2XH 100 pH Parameter Channel X Sub parameter a b c d Temperature soure (analog sensor) pH buffer Drift Control IP STC Fix CalTemp Cal constants (for Analog sensor) Cal constants (for ISM sensor) Resolution Analog outputs pH temperature Set point 1 Set point 2 Alarm pH temperature pH Temperature 1 2 Value 4 mA value 20 mA Value 4 mA value 20 mA measurement type OC measurement Type OC Rg diagnostics Rr diagnostics © 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switz
Transmitter M400/2(X)H, M400G/2XH 101 pH/pNa Parameter Channel X Temperature soure (analog sensor) pH buffer Drift Control IP STC Fix CalTemp Cal constants Resolution Analog outputs pH temperature Set point 1 Set point 2 Alarm Sub parameter a b c d pH Temperature 1 2 Value 4 mA value 20 mA Value 4 mA value 20 mA measurement type OC measurement Type OC Rg diagnostics © 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland Value pH temperature None None Auto Na+3.
Transmitter M400/2(X)H, M400G/2XH 102 Oxygen Parameter Sub parameter Value Channel X a O2 b temperature c O2(dual channel) d Unit %Air (O2 low:ppb) °C %Air (O2 low:ppb) Analog outputs Temperature 1 2 temperature(dual channel) Auto 759.8 759.8 CaPres Auto 0.0 100 Read form sensor -674 S=-70.00nA,Z=0.00nA S=-350.00nA,Z=0.00nA M=1.0, A=0.0 Read from sensor 0.1 1 0.
Transmitter M400/2(X)H, M400G/2XH 103 Resistivity/Conductivity Parameter Channel X Sub parameter a b c d Temperature soure (analog sensor) Compension Cal constants (for analog sensor) Cal constants (for ISM sensor) Resolution Analog outputs Conductivity/Resistivity Temperature Set point 1 Set point 2 Alarm Cond/Res temperature Resisitivity Temperature 1 2 Value 4 mA Value 20 mA value 4 mA value 20 mA measurement type OC measurement Type OC Cond cell shorted Dry cond sensor Cell deviation (ISM senso
Transmitter M400/2(X)H, M400G/2XH 18 104 Warranty METTLER TOLEDO warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. METTLER TOLEDO’s Customer Service Dept. will determine if the product problem is due to deviations or customer abuse.
Transmitter M400/2(X)H, M400G/2XH 19 105 Buffer tables M400 transmitters have the ability to do automatic pH buffer recognition. The following tables show different standard buffers that are automatically recognized. 19.1 Standard pH buffers 19.1.1 Mettler-9 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 pH of buffer solutions 2.03 4.01 2.02 4.01 2.01 4.00 2.00 4.00 2.00 4.00 2.00 4.01 1.99 4.01 1.99 4.02 1.98 4.03 1.98 4.04 1.98 4.06 1.98 4.08 1.98 4.10 1.98 4.13 1.99 4.
Transmitter M400/2(X)H, M400G/2XH 19.1.2 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 19.1.3 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 106 Mettler-10 pH of buffer solutions 2.03 4.01 2.02 4.01 2.01 4.00 2.00 4.00 2.00 4.00 2.00 4.01 1.99 4.01 1.99 4.02 1.98 4.03 1.98 4.04 1.98 4.06 1.98 4.08 1.98 4.10 1.99 4.13 1.98 4.16 1.99 4.19 2.00 4.22 2.00 4.26 2.00 4.30 2.00 4.35 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.
Transmitter M400/2(X)H, M400G/2XH 19.1.4 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 70 80 90 95 h 107 NIST standard buffers (DIN and JIS 19266: 2000–01) pH of buffer solutions 1.668 1.670 1.672 1.676 1.680 1.685 1.694 1.697 1.704 1.712 1.715 1.723 1.743 1.766 1.792 1.806 4.004 4.001 4.001 4.003 4.008 4.015 4.028 4.036 4.049 4.064 4.075 4.091 4.126 4.164 4.205 4.227 6.950 6.922 6.900 6.880 6.865 6.853 6.841 6.837 6.834 6.833 6.834 6.836 6.845 6.859 6.877 6.886 9.392 9.331 9.277 9.228 9.184 9.
Transmitter M400/2(X)H, M400G/2XH 19.1.6 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 108 Ciba (94) buffers pH of buffer solutions 2.04 4.00 2.09 4.02 2.07 4.00 2.08 4.00 2.09 4.01 2.08 4.02 2.06 4.00 2.06 4.01 2.07 4.02 2.06 4.03 2.06 4.04 2.05 4.05 2.08 4.10 2.07* 4.10* 2.07 4.11 2.04* 4.13* 2.02 4.15 2.03* 4.17* 2.04 4.20 2.05* 4.22* 7.10 7.08 7.05 7.02 6.98 6.98 6.96 6.95 6.94 6.93 6.93 6.91 6.93 6.92* 6.92 6.92* 6.93 6.95* 6.97 6.99* 10.30 10.21 10.14 10.06 9.99 9.95 9.
Transmitter M400/2(X)H, M400G/2XH 19.1.8 Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 19.1.9 Temp (°C) 0 5 10 15 20 25 30 35 38 40 45 50 55 60 70 80 90 95 109 WTW buffers pH of buffer solutions 2.03 4.01 2.02 4.01 2.01 4.00 2.00 4.00 2.00 4.00 2.00 4.01 1.99 4.01 1.99 4.02 1.98 4.03 1.98 4.04 1.98 4.06 1.98 4.08 1.98 4.10 1.99 4.13 4.16 4.19 4.22 4.26 4.30 4.35 7.12 7.09 7.06 7.04 7.02 7.00 6.99 6.98 6.97 6.97 6.97 6.98 6.98 6.99 7.00 7.02 7.04 7.06 7.09 7.12 10.65 10.52 10.
Transmitter M400/2(X)H, M400G/2XH 110 19.2 Dual membrane pH electrode buffers 19.2.1 Mettler-pH/pNa buffers (Na+ 3.9M) Temp (°C) 0 5 10 15 20 25 30 35 40 45 50 pH of buffer solutions 1.98 3.99 1.98 3.99 1.99 3.99 1.99 3.99 1.99 4.00 2.00 4.01 2.00 4.02 2.01 4.04 2.01 4.05 2.02 4.07 2.02 4.09 © 03 / 2014 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Printed in Switzerland 7.01 7.00 7.00 6.99 7.00 7.00 7.01 7.01 7.02 7.03 7.04 9.51 9.43 9.36 9.30 9.25 9.21 9.18 9.15 9.12 9.11 9.
METTLER TOLEDO Market Organizations Sales and Service: Australia Mettler-Toledo Ltd. 220 Turner Street Port Melbourne AUS - 3207 Melbourne / VIC Phone +61 1300 659 761 Fax +61 3 9645 3935 e-mail info.mtaus@mt.com France Mettler-Toledo Analyse Industrielle S.A.S. 30, Boulevard de Douaumont F - 75017 Paris Phone +33 1 47 37 06 00 Fax +33 1 47 37 46 26 e-mail mtpro-f@mt.com Austria Mettler-Toledo Ges.m.b.H. Südrandstraße 17 A -1230 Wien Phone +43 1 604 19 80 Fax +43 1 604 28 80 e-mail infoprocess.mtat@mt.