Specifications

Transmitter Multiparameter M400 47
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
The sensor keeps the information stored in the built in electronics and can be retrieved via a
transmitter or the ISense asset management suite.
The alarm will be reseted if the Lifetime Indicator is not 0 days anymore (e.g. after connecting a
new sensor or changes on the measurement conditions).
For amperometric Oxygen sensors, the lifetime indicator is related to the inner-body of the
sensor. After exchanging the inner-body, reset the lifetime indicator in the ISM Setup menu (menu
Configure).
For optical DO sensors the lifetime indicator is related to the OptoCap. After exchanging the
OptoCap, reset the lifetime indicator in the ISM Setup menu (menu Configure).
If the Lifetime Indicator is turned on, the value will be automatically shown in the display on line 3.
Adaptive Cal Timer: This timer estimates when the next calibration should be performed to keep
the best possible measurement performance. The timer is influenced by significant changes on
the DLI parameters.
Adaptive Cal Timer YES / NO
Alarm YES / NO R# choose relay
The alarm will be reseted after a successful calibration.
If the Adaptive Cal Timer is turned on, the value will be automatically shown in the display on
line 4.
C
NOTE: The recommended initial value for the calibration interval will be uploaded from the
sensor to the transmitter and can be adapted according to the application experience (menu
Configure / ISM Setup / Sensor monitoring).
Time to Maintenance (not for optical sensors): This timer estimates when the next cleaning
cycle should be performed to keep the best possible measurement performance. The timer is
influenced by significant changes on the DLI parameters.
Time to Maintenance YES / NO
Alarm YES / NO R# choose relay
The alarm needs to be reseted in the menu ”ISM Setup”. For oxygen sensors, the time to
maintenance indicates a maintenance cycle for the membrane and electrolyte.
C
NOTE: The recommended initial value for the maintenance interval will be uploaded from the
sensor to the transmitter and can be adapted according to the application experience (menu
Configure / ISM Setup / Sensor monitoring).
8.6.2 CIP Cycle Limit
Please activate the function ISM (see chapter 8.2.1) configuring the CIP cycle limit.
The CIP cycle limit counts the number of CIP cycles if the limit (user defined) is reached, an
alarm can be indicated and set to a certain output relays. The following options are possible:
CIP Max 000
Alarm YES / NO R# choose relay
If the Max setting is on 000, the counter functionality is turned off. The alarm will be reseted after
exchanging the sensor. For oxygen sensors, the counter can be reseted in the ISM Setup menu
(menu Configure)