Specifications

Table Of Contents
Transmitter M400 95
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
Pulse Frequency – If using a pulse input metering pump, select ”Relays” and ”PF”, pulse fre-
quency. Choose the first relay position as #3 and / or the second relay position as #4 according
to the table below. Set the pulse frequency to the maximum frequency allowed for the particular
pump being used, typically 60 to 100 pulses / minute. Control action will produce this frequency
at 100% output.
h
NOTE: All relays from #1 to #6 can be used for the controlling function.
CAUTION: Setting the pulse frequency too high may cause the pump to overheat.
1
st
Relay Position = #3 2
nd
Relay Position = #4 Pulse Frequency (PF)
Conductivity
Controlling
concentrating
chemical feed
Controlling dilution water
Max allowed for the
pump used (typically
60–100
pulses / minute)
pH / ORP Feeding base Feeding acid
Max allowed for the
pump used (typically
60–100
pulses / minute)
Dissolved
Oxygen
Reverse control action
Direct acting control
action
Max allowed for the
pump used (typically
60–100
pulses / minute)
Analog – If using analog control, change ”Relays” to ”Analogout” using up / down arrow keys.
Choose the first Analogout position as #1 and / or the second Analogout position as #2 accord-
ing to the table below. Select the analog output current range required by the control device,
4–20 or 0–20 mA. Press [ENTER].
1
st
Analogout Position = #1 2
nd
Analogout Position = #2
Conductivity
Controlling concentrating
chemical feed
Controlling dilution water
pH / ORP Feeding base Feeding acid
Dissolved Oxygen Reverse control action Direct acting control action
10.4 Tune parameters
(PATH: MENU / PID Setup / Tune Parameters)
This menu assigns control to a measurement and sets the setpoint, tuning parameters and non-
linear functions of the controller through a series of screens.