Operating instructions
Settings in the master mode
16 Operating instructions 22012836B 09/08
IND690-Fill
AUTOMATIC TARE Switch automatic taring before dispensing on or off
Factory setting: AUTOMATIC TARE ON
REDISP. CORRECTION Switch redispensing correction on or off
The redispensing correction function optimizes the switch-off point of the fine feed
(limit 2).
If REDISP. CORRECTION ON is set, the target-actual difference is determined for
each container and multiplied by a FACTOR.
Target-actual difference x correction factor = Δ
Limit 2 is automatically corrected by the value Δ when dispensing the next
container:
Example: For a target-actual difference of 10 g and a factor of 0.5, limit 2 is
corrected by 5 g.
Factory setting: REDISP. CORRECTION ON
FACTOR Correction factor by which the target-actual difference is multiplied. The result is the
value
Δ by which limit 2 is corrected.
Possible values: 0.1 … 0.9 (factory setting: 0.5)
CORREC. THRESHOLD The correction threshold specifies the target-actual difference up to which the
redispensing correction corrects limit 2.
• Possible values: O … 99 in multiples of the tolerance
(Factory setting: 0, i. e. limit 2 is corrected for all actual values)
• Limit 2 is not corrected when TOTALIZING ON is set and after at least 10
consecutive dispensings the actual value lies outside the correction threshold for
the first time. This value is considered a freak value. If during the next dispensing
the actual value lies outside the correction threshold, limit
2 is automatically
corrected. If in the process limit 2
≤ limit 1, then the learn mode is automatically
activated.