Operating instructions

Settings in the master mode
Operating instructions 22012836B 09/08 17
IND690-Fill
FIXED TARGET Save dispensing parameters for various components in target memories
protected against power failure
LOAD FIXED TARGET 1. Enter FIXED TARGET NO. memory number: 1 … 999.
2. Enter article designation NAME, e.g. M8 SCREW.
3. Enter TARG target weight.
4. If OUTPUT 2 = 3RD SHUTOFF POINT: Enter switchover point preflow/coarse feed
LIMIT
0.
5. Enter coarse/fine feed switchover point LIM 1.
6. Enter switch-off point of fine feed LIM 2: LIMIT 1 LIMIT 2.
7. Enter tolerance TOL in the displayed unit.
Minimum tolerance: 1 digit
Maximum tolerance: target weight; with DeltaTrac: 10 % of target weight
Target weight + tolerance maximum load
8. Enter lower limit of permissible tare range TMIN.
9. Enter upper limit of permissible tare range TMAX: TMIN TMAX.
10. If ANALOG OUTPUT = ON and OUTPUT 2 = 3rd SHUTOFF POINT:
Enter THROUGHPUT PREFLOW.
If ANALOG OUTPUT = ON:
Enter THROUGHPUT COARSE FEED and THROUGHPUT FINE FEED
11. End entry: Confirm memory number without entry with ENTER.
CLEAR ALL TARGET V. Clear all target memories.
REDISPENSING Set automatic or manual redispensing
Factory setting: AUTOMAT. REDISPENSING
AUTOMAT.
REDISPENSING
MANUAL
REDISPENSING
Possible entries:
PULSE DURATION
During the pulse duration the fine feed is opened.
Possible values: 1 … 99 times a measuring cycle (factory setting: 5)
PULSE PAUSE
During the pulse pause the fine feed is closed.
Possible values: 0 … 99 times a measuring cycle (factory setting: 5)