Operating instructions
Settings in the master mode
Operating instructions 22012836B 09/08 19
IND690-Fill
OPERATING MODE Set operating mode with certified weighing platforms
NON AUTOMATIC The dispensing process does not run automatically and the permissibility of the
weight values must be monitored by the operator.
AUTOMATIC The dispensing process runs automatically (factory setting).
Comments • Different national tolerances are taken into account.
• For calibration reasons, the operating mode can only be switched over in the
non-certified mode of the weighing platform.
MANUAL CORRECTION Switch manual recorrection on or off
When MAN. CORRECTION ON is set, the final weight can be manually recorrected,
e.g. in the case of incorrect dispensing, see section
1.10.
Factory setting: MAN. CORRECTION OFF
Comments • At output OUT4 and OUT5 of the first 8-690 relay box, it can be read off whether
dispensing lies within the tolerances (DISPENSING OKAY) or outside (DISPENS
-
ING POOR).
• Manual correction is not possible in case of verifiable weighing platforms in the
AUTOMATIC operating mode.
• If REDISPENSING is set to AUTOMATIC, the MANUAL CORRECTION only becomes
active in the case of overfilling (underfilled containers are automatically
redispensed). If REDISPENSING is set to MANUAL, the MANUAL CORRECTION
becomes active in the case of underfilling and overfilling.
OUTPUT 2 Control various additional devices via output 2
OUTPUT 2 OFF Output 2 is not actuated (factory setting).
MATERIAL AGITATION Control of an agitator during or after dispensing;
for additional settings, see section 2.2.1.
REMAINING QUANTITY Remaining quantity: Control of an emptying device on the filling container;
for additional settings, see section 2.2.2.
FILL QUANTITY Fill quantity: Control of a refilling valve during subtractive weighing;
for additional settings, see section 2.2.3.
SCALE SWITCHOVER Control of a signal which enables switching over between 2 weighing platforms;
for additional settings, see section 2.2.4.
3RD SHUTOFF POINT Control of a third valve; for additional settings, see section 2.2.5.
Comment To read or set the status of output 2, see application block 359 on page 35.